A commercial version of a synthetic single crystal diamond detector (SCDD) in a Schottky diode configuration was recently released as the new type 60019 microDiamond detector (PTW-Freiburg, Germany). In this study we investigate the dosimetric properties of this detector to independently confirm that findings from the developing group of the SCDDs still hold true for the commercial version of the SCDDs. We further explore if the use of the microDiamond detector can be expanded to high-energy photon beams of up to 15 MV and to large field measure- ments. Measurements were performed with an Elekta Synergy linear accelerator delivering 6, 10, and 15 MV X-rays, as well as 6, 9, 12, 15, and 20 MeV electron beams. The dependence of the microdiamond detector response on absorbed dose after connecting the detector was investigated. Furthermore, the dark current of the diamond detector was observed after irradiation. Results are compared to similar results from measurements with a diamond detector type 60003. Energy dependency was investigated, as well. Photon depth-dose curves were measured for field sizes 3 × 3, 10 × 10, and 30 × 30 cm2. PDDs were measured with the Semiflex type 31010 detector, microLion type 31018 detector, P Diode type 60016, SRS Diode type 60018, and the microDiamond type 60019 detector (all PTW-Freiburg). Photon profiles were measured at a depth of 10 cm. Electron depth-dose curves normalized to the dose maximum were measured with the 14 × 14 cm2 electron cone. PDDs were measured with a Markus chamber type 23343, an E Diode type 60017 and the microDiamond type 60019 detector (all PTW-Freiburg). Profiles were measured with the E Diode and microDiamond at half of D90, D90, D70, and D50 depths and for electron cone sizes of 6 × 6 cm2, 14 × 14 cm2, and 20 × 20 cm2. Within a tol- erance of 0.5% detector response of the investigated detector was stable without any preirradiation. After preirradition with approximately 250 cGy the detector response was stable within 0.1%. A dark current after irradiation was not observed. The microDiamond detector shows no energy dependence in high energy photon or electron dosimetry. Electron PDD measurements with the E Diode and micro- Diamond are in good agreement. However, compared to E Diode measurements, dose values in the bremsstrahlungs region are about 0.5% lower when measured with the microDiamond detector. Markus detector measurements agree with E Diode measurements in the bremsstrahlungs region. For depths larger than dmax, depth-dose curves of photon beams measured with the microDiamond detector are in close agreement to those measured with the microLion detector for small fields and with those measured with a Semiflex 0.125 cc ionization chamber for large fields. Differences are in the range of 0.25% and less. For profile measurements, microDiamond detector measurements agree well with microLion and P Diode measurements in the high-dose region of the profile and the penumbra region. For areas outside the open field, P Diode measurements are about 0.5%-1.0% higher than microDiamond and microLion measurements. Thus it becomes evident that the investigated diamond detector (type 60019) is suitable for a wide range of applications in high-energy photon and electron dosimetry and is interesting for relative, as well as absolute, dosimetry. 

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5711122PMC
http://dx.doi.org/10.1120/jacmp.v15i6.4890DOI Listing

Publication Analysis

Top Keywords

microdiamond detector
28
detector
23
diamond detector
20
type 60019
16
diode measurements
16
measurements
12
detector ptw-freiburg
12
detector response
12
depth-dose curves
12
diode type
12

Similar Publications

Article Synopsis
  • Proton Minibeam Radiation Therapy has been promising in enhancing treatment efficacy compared to traditional radiation, but more research into its biological mechanisms is needed.
  • A mechanical collimation setup was developed to produce 250µm minibeams with a 1000µm spacing, with optimization using Monte Carlo simulations conducted at various proton therapy sites.
  • Results showed a peak-to-valley dose ratio (PVDR) of 10 in Dresden and 14 in Seattle, with some discrepancies between dosimetry methods that can be addressed with correction factors.
View Article and Find Full Text PDF

. To investigate the effect of the position and orientation of the detector and its influence on the determination of output factors (OF) for small fields for a linear accelerator (MR-linac) integrated with 1.5 T magnetic resonance following the TRS-483 formalism.

View Article and Find Full Text PDF

Purpose: To provide beam quality correction factors ( ) for detectors used in Ir brachytherapy dosimetry measurements.

Materials And Methods: Ten detectors were studied, including the PTW 30013 and Exrading12 Farmer large cavity chambers, seven medium (0.1-0.

View Article and Find Full Text PDF

Dosimetric study of bevel factors in IOERT with mobile linacs: Towards a unified code of practice.

Phys Med

November 2024

Nuclear Physics Group and IPARCOS, Department of Structure of Matter, Thermal Physics and Electronics, CEI Moncloa, Universidad Complutense de Madrid, Madrid, Spain; Instituto de Investigación Sanitaria del Hospital Clínico San Carlos, Madrid, Spain.

Background: Dosimetry in intraoperative electron radiotherapy (IOERT) poses distinct challenges, especially with inclined applicators deviating from international protocols. Ion recombination in ionization chambers, electron beam degradation due to scattering in cylindrical applicators, coupled with a lack of a well-defined beam quality surrogate, complicate output factor determination with ionization chambers. Synthetic diamond-based detectors, offer potential solutions; however, their suitability requires further exploration.

View Article and Find Full Text PDF

A review of diamond dosimeters in advanced radiotherapy techniques.

Med Phys

December 2024

Radiotherapy and Radiation Dosimetry, National Physical Laboratory (NPL), Teddington, UK.

This review article synthesizes key findings from studies on the use of diamond dosimeters in advanced radiotherapy techniques, showcasing their applications, challenges, and contributions to enhancing dosimetric accuracy. The article explores various dosimeters, highlighting synthetic diamond dosimeters as potential candidates especially due to their high spatial resolution and negligible ion recombination effect. The clinically validated commercial dosimeter, PTW microDiamond (mD), faces limitations in small fields, proton and hadron therapy and ultra-high dose per pulse (UHDPP) conditions.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!