The Gown-glove Interface Is a Source of Contamination: A Comparative Study.

Clin Orthop Relat Res

Department of Orthopaedic Surgery, Banner Good Samaritan Hospital, 1320 N 10th Street, Suite A, Phoenix, AZ, 85006, USA,

Published: July 2015

Background: The original Charnley-type negative-pressure body exhaust suit reduced infection rates in randomized trials of total joint arthroplasty (TJA) decades ago. However, modern positive-pressure surgical helmet systems have not shown similar benefit, and several recent studies have raised the question of whether these gowning systems result in increased wound contamination and infections. The gown-glove interface may be one source of particle contamination.

Questions/purposes: The purpose of this study was to compare particle contamination at the gown-glove interface in several modern surgical helmet systems and conventional surgical gowns.

Methods: A 5-μm fluorescent powder was evenly applied to both hands to the level of the wrist flexion crease. After gowning in the standard fashion, the acting surgeon performed a 20-minute simulated TJA protocol. Each of the five gowning systems was run through five trials. The amount of gown contamination at the gown-glove interface then was measured by three observers under ultraviolet light using a grading scale from 0 (no contamination) to 4 (gross contamination). Statistical analysis was carried out with Minitab 15. Friedman's test was used to compare the levels of contamination across trials for each gown and the Mann-Whitney test was used post hoc to perform a pairwise comparison of each gown.

Results: All gown-glove interfaces showed some contamination. Friedman's test showed that there was a significant difference in contamination between gowns (p = 0.029). The Stryker T5 Zipper Toga system showed more contamination than the other gowns. The median contamination score and range for each gowning setup was 1.8 (range, 1-4; conventional Kimberly-Clark MicroCool gown without helmet), 4 (range, 3-4; Stryker T5 Zipper Toga), 3.6 (range, 0-4; Stryker helmet with conventional gown), 1.6 (range, 0-2; Stryker Flyte Toga), and 3.0 (range, 2-3; DePuy Toga). A Mann-Whitney test found no difference among any of the gowns except for the Stryker T5 Zipper Toga, which showed more contamination compared directly with each of the other four gowns (p < 0.001 for each gown-to-gown comparison).

Conclusions: Particle contamination occurs at the gown-glove interface in most commonly used positive-pressure surgical helmet systems. The Stryker T5 Zipper Toga exhibited more contamination than each of the other gowning systems.

Clinical Relevance: The gown-glove interface is prone to particle contamination and all surgeons should be aware of this area as a potential source of surgical site infection. Although future studies are needed to clarify the link between particle contamination through this route and clinical infection, surgeons should consider using gowning systems that minimize the migration of fomites through the gown-glove interface.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4457760PMC
http://dx.doi.org/10.1007/s11999-014-4094-8DOI Listing

Publication Analysis

Top Keywords

gown-glove interface
28
contamination
16
particle contamination
16
stryker zipper
16
zipper toga
16
surgical helmet
12
helmet systems
12
gowning systems
12
gown-glove
8
interface source
8

Similar Publications

Donning Gloves Before Surgical Gown Cross-contaminates the Assistant.

Arthroplast Today

October 2022

Department of Orthopaedics and Rehabilitation, University of Vermont College of Medicine, Burlington, VT, USA.

Background: Maintaining a sterile surgical environment is of the upmost importance in total joint arthroplasty. The surgical gown-glove interface is a frequent and known cause of bacterial contamination. A variation of the traditional staff-assisted open-gloving technique has been developed in an attempt to address this.

View Article and Find Full Text PDF

The sterility of the gown-glove interface during total joint arthroplasty is a key factor in preventing contamination of the surgical field. To compare the potential of gown-glove interface contamination with a novel gloving technique versus standard gloving technique. We performed a study quantifying potential gown-glove interface contamination using two different gloving techniques.

View Article and Find Full Text PDF

Background: Surgical helmet systems commonly are stand-alone systems with a single fan blowing air into the suit, creating positive pressure that blows particles out through areas of low resistance, possibly contaminating surgical attire and the surgical field. Two-fan systems were developed more recently to release spent air, also theoretically lowering pressure in the suit and decreasing the aforementioned risk of particle contamination. To our knowledge no study to date has measured the potential differences in gown particle contamination to support this hypothesis.

View Article and Find Full Text PDF

Do surgical helmet systems affect intraoperative wound contamination? A randomised controlled trial.

Arch Orthop Trauma Surg

November 2017

Department of Orthopaedic Surgery, North Shore Hospital, University of Auckland, 124 Shakespeare Road, Takapuna Private Bag 93-503, Auckland, 0740, New Zealand.

Background: Deep infection following arthroplasty remains a devastating complication. Some registry data suggests that modern positive-pressure surgical helmet systems (SHS) are associated with a paradoxical increase in infection rates, and as such their role in arthroplasty remains unclear. The aim of this study was to investigate whether SHS increase wound contamination in total knee arthroplasty (TKA) and if this contamination can be reduced by placing tape around the gown/glove interface.

View Article and Find Full Text PDF

Surgical Gown's Cuff Modification to Prevent Surgical Contamination.

J Maxillofac Oral Surg

June 2015

Oral and Maxillofacial Surgery Department, University Hospital La Paz, Pº de la Castellana 261, 28046 Madrid, Spain.

The gown-glove interface is the weakest point in the barrier system of gown and glove protection for the surgeon. We are herein presenting an easy modification in the cuff of the gown that increases the security of the gown-glove interface.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!