Background: The use of dual growing rods is a fusionless surgical approach to the treatment of early onset scoliosis which aims to harness potential growth and correct spinal deformity. The purpose of this study was to compare the in-vitro biomechanical response of two different dual rod designs under axial rotation loading.
Methods: Six porcine spines were dissected into seven level thoracolumbar multi-segment units. Each specimen was mounted and tested in a biaxial Instron machine, undergoing nondestructive left and right axial rotation to peak moments of 4 Nm at a constant rotation rate of 8 deg. s(-1). A motion tracking system (Optotrak) measured 3D displacements of individual vertebrae. Each spine was tested in an un-instrumented state first and then with appropriately sized semi-constrained and 'rigid' growing rods in alternating sequence. The range of motion, neutral zone size and stiffness were calculated from the moment-rotation curves and intervertebral range of motion was calculated from Optotrak data.
Findings: Irrespective of test sequence, rigid rods showed a significant reduction of total rotation across all instrumented levels (with increased stiffness) whilst semi-constrained rods exhibited similar rotational behavior to the un-instrumented spines (P<0.05). An 11.1% and 8.0% increase in stiffness for left and right axial rotation respectively and 14.9% reduction in total range of motion were recorded with dual rigid rods compared with semi-constrained rods.
Interpretation: Based on these findings, the Semi-constrained growing rods were shown to not increase axial rotation stiffness compared with un-instrumented spines. This is thought to provide a more physiological environment for the growing spine compared to dual rigid rod constructs.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.clinbiomech.2014.11.008 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!