Relationship between transient and steady-state pattern electroretinograms: theoretical and experimental assessment.

Invest Ophthalmol Vis Sci

University of Miami, Department of Biomedical Engineering, Coral Gables, Florida, United States University of Miami, Bascom Palmer Eye Institute, Miami, Florida, United States.

Published: December 2014

Purpose: We determined if the overlap of transient (tr) pattern electroretinograms (PERG(tr)) can explain the generation of the steady-state (SS) pattern electroretinogram (PERG(SS)), and investigated the relationship between the two types of responses.

Methods: Slightly jittered pattern reversals were used to generate quasi SS (QSS) PERG(SS) responses from eight normal subjects, recorded using lower eyelid skin electrodes, at rates between 6.9 and 26.5 reversals per second (rps). Jittered quasi PERG(SS) were deconvolved using the frequency domain continuous loop averaging deconvolution method. Additionally, conventional PERG(tr) at 2.2 rps and PERG(SS) at each of the QSS stimulation rates were obtained from all subjects. Two synthetic PERG(SS) responses were constructed at each stimulation rate, one using the PERG(tr) obtained at that rate, and the other using the conventional 2.2 rps PERG(tr). Synthetic responses then were compared to the recorded PERG(SS) using amplitude, latency, and spectral measurements.

Results: Findings indicate that the PERG(SS) obtained at SS rates can be predicted using the superposition of deconvolved tr PERGs at each particular rate. Although conventional PERG(tr) can explain PERG(SS) obtained at rates below 15.4 rps (≥ 97% correlation), for higher reversal rates only deconvolved responses obtained at that rate can produce the recorded SS responses (96% vs. 65% correlation at 26.5 rps).

Conclusions: The study shows that PERG(SS) results from the overlapping of tr PERG(tr) waveforms generated at that reversal rate. The first two peaks (N(SS) and P(SS)) of the PERG(SS) reflect N35 and P50 waves of the tr PERG(tr). The N95 amplitude is reduced at conventional (16 rps) SS rates, but contributes to the overall PERG(SS) amplitude.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4280086PMC
http://dx.doi.org/10.1167/iovs.14-15685DOI Listing

Publication Analysis

Top Keywords

pergss
11
steady-state pattern
8
pattern electroretinograms
8
pergtr explain
8
pergss responses
8
conventional pergtr
8
rate conventional
8
conventional rps
8
pergss amplitude
8
pergss rates
8

Similar Publications

Relationship between transient and steady-state pattern electroretinograms: theoretical and experimental assessment.

Invest Ophthalmol Vis Sci

December 2014

University of Miami, Department of Biomedical Engineering, Coral Gables, Florida, United States University of Miami, Bascom Palmer Eye Institute, Miami, Florida, United States.

Purpose: We determined if the overlap of transient (tr) pattern electroretinograms (PERG(tr)) can explain the generation of the steady-state (SS) pattern electroretinogram (PERG(SS)), and investigated the relationship between the two types of responses.

Methods: Slightly jittered pattern reversals were used to generate quasi SS (QSS) PERG(SS) responses from eight normal subjects, recorded using lower eyelid skin electrodes, at rates between 6.9 and 26.

View Article and Find Full Text PDF

Unwrapping of transient responses from high rate overlapping pattern electroretinograms by deconvolution.

Clin Neurophysiol

October 2014

University of Miami, Dept. Biomedical Engineering, PO Box 248294, Coral Gables, FL 33124, United States; University of Miami, Depts. Otolaryngology, Pediatrics and Neuroscience (Graduate), United States. Electronic address:

Objective: Due to overlapping, temporal information is mostly lost in high rate steady-state pattern electroretinograms (PERGSS). This study develops a deconvolution method and a display/recording system to "unwrap" PERGSS and obtain a transient, "per stimulus" response (PERGtr) regardless of reversal rate.

Methods: Processing and instrumentation, including high temporal resolution display and acquisition were developed for deconvolving PERGs acquired at high rates by slight jittering of reversal onsets at a given mean rate.