A single-step high performance liquid chromatography (HPLC) method for vitamin D quantitation in blood serum is described. Methanol/hexane extraction, silica Sep-Pak purification, differential dissolution, filtration, and the use of a precolumn in place of an injector sample loop allow for using only one (reversed phase) HPLC step instead of the usual two. Detection limit with the HPLC system used was about 1 micrograms/L of serum. Recovery of vitamin D3 added to serum ranged from 90% to 109%. Intra and interassay coefficient of variation values were 8.1% and 15.4%, respectively.

Download full-text PDF

Source
http://dx.doi.org/10.1016/0160-5402(89)90062-4DOI Listing

Publication Analysis

Top Keywords

hplc method
8
blood serum
8
simplified hplc
4
method quantitation
4
quantitation vitamin
4
vitamin blood
4
serum
4
serum single-step
4
single-step high
4
high performance
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!