Optical simulation for subsurface nanoglistening.

J Cataract Refract Surg

From the Department of Ophthalmology (Takahashi, Mita, Hatsusaka, S. Shibata, N. Shibata, Kubo, Sasaki) and the Division of Vision, Research for Environmental Health (Hatsusaka, Sasaki), Kanazawa Medical University, Uchinada, the Department of Orthoptics and Visual Science, Kitasato University School of Allied Health Sciences (Kawamorita), and the Department of Ophthalmology, Kitasato University School of Medicine (Kawamorita), Sagamihara, Japan.

Published: January 2015

AI Article Synopsis

  • The study aimed to assess whether subsurface nanoglistening in hydrophobic acrylic intraocular lenses affects visual performance.
  • The researchers simulated the impact of nanoglistening using optical design software and evaluated various parameters, including retinal irradiance and light scattering, with different particle sizes and volume ratios.
  • Results indicated that while subsurface nanoglistening led to slight increases in light scattering and reduced peak retinal irradiance, the overall effect on visual function was minimal in the absence of severe retinal issues.

Article Abstract

Purpose: To determine whether subsurface nanoglistening in hydrophobic acrylic intraocular lenses (IOL) diminishes visual performance.

Setting: Department of Ophthalmology, Kanazawa Medical University, Uchinada, Ishikawa, Japan.

Design: Experimental study.

Methods: The effect of subsurface nanoglistenings was simulated using optical design software Lighttools and Code V with the Liou-Brenann model eye and an acrylic IOL. Peak irradiance of the retina, forward light scattering, and modulation transfer function (MTF) were evaluated. During optical simulation, particle diameters were set at 100 nm, 150 nm, and 200 nm and volume ratios at 0%, 0.05%, 0.1%, 0.2%, 0.5%, and 1.0%.

Results: Peak irradiance decreased as subsurface nanoglistening volume ratio and particle size increased. At a volume ratio of 0.05%, the peak irradiance of subsurface nanoglistening particles 100 nm, 150 nm, and 200 nm in diameter decreased 0.7%, 1.8%, and 2.9%, respectively, compared with those at volume ratio 0% (no subsurface nanoglistenings). At a volume ratio of 0.1%, the peak irradiance of subsurface nanoglistening particles 100 nm, 150 nm, and 200 nm decreased 1.5%, 3.6%, and 5.7%, respectively. Forward light scattering increased with increased size of subsurface nanoglistening particle and volume ratio. The MTF was not altered by changes in subsurface nanoglistening particle size or volume ratio.

Conclusions: Subsurface nanoglistenings increased forward scattering slightly and reduced irradiance but significantly diminished retinal image. The effect of subsurface nanoglistenings on visual function in the absence of severe retinal disease was minimal.

Financial Disclosure: No author has a financial or proprietary interest in any material or method mentioned.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jcrs.2014.10.021DOI Listing

Publication Analysis

Top Keywords

subsurface nanoglistening
28
volume ratio
20
subsurface nanoglistenings
16
peak irradiance
16
100 150
12
150 200
12
subsurface
11
optical simulation
8
forward light
8
light scattering
8

Similar Publications

Background: To report on five patients with decreased visual acuity due to glistening and severe sub-surface nano-glistening (SSNG) formation within their intraocular lenses (IOLs).

Design: Case reports and analysis of extracted IOLs.

Participants And Samples: We report improved visual acuity when IOLs with severe glistening and SSNG were exchanged for clear IOLs in five patients.

View Article and Find Full Text PDF

Optical simulation for subsurface nanoglistening.

J Cataract Refract Surg

January 2015

From the Department of Ophthalmology (Takahashi, Mita, Hatsusaka, S. Shibata, N. Shibata, Kubo, Sasaki) and the Division of Vision, Research for Environmental Health (Hatsusaka, Sasaki), Kanazawa Medical University, Uchinada, the Department of Orthoptics and Visual Science, Kitasato University School of Allied Health Sciences (Kawamorita), and the Department of Ophthalmology, Kitasato University School of Medicine (Kawamorita), Sagamihara, Japan.

Article Synopsis
  • The study aimed to assess whether subsurface nanoglistening in hydrophobic acrylic intraocular lenses affects visual performance.
  • The researchers simulated the impact of nanoglistening using optical design software and evaluated various parameters, including retinal irradiance and light scattering, with different particle sizes and volume ratios.
  • Results indicated that while subsurface nanoglistening led to slight increases in light scattering and reduced peak retinal irradiance, the overall effect on visual function was minimal in the absence of severe retinal issues.
View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!