Background: Limited information exists describing the results of transcatheter aortic valve (TAV) replacement in patients with bicuspid aortic valve (BAV) disease (TAV-in-BAV).

Objectives: This study sought to evaluate clinical outcomes of a large cohort of patients undergoing TAV-in-BAV.

Methods: We retrospectively collected baseline characteristics, procedural data, and clinical follow-up findings from 12 centers in Europe and Canada that had performed TAV-in-BAV.

Results: A total of 139 patients underwent TAV-in-BAV with the balloon-expandable transcatheter heart valve (THV) (n = 48) or self-expandable THV (n = 91) systems. Patient mean age and Society of Thoracic Surgeons predicted risk of mortality scores were 78.0 ± 8.9 years and 4.9 ± 3.4%, respectively. BAV stenosis occurred in 65.5%, regurgitation in 0.7%, and mixed disease in 33.8% of patients. Incidence of type 0 BAV was 26.7%; type 1 BAV was 68.3%; and type 2 BAV was 5.0%. Multislice computed tomography (MSCT)-based TAV sizing was used in 63.5% of patients (77.1% balloon-expandable THV vs. 56.0% self-expandable THV, p = 0.02). Procedural mortality was 3.6%, with TAV embolization in 2.2% and conversion to surgery in 2.2%. The mean aortic gradient decreased from 48.7 ± 16.5 mm Hg to 11.4 ± 9.9 mm Hg (p < 0.0001). Post-implantation aortic regurgitation (AR) grade ≥ 2 occurred in 28.4% (19.6% balloon-expandable THV vs. 32.2% self-expandable THV, p = 0.11) but was prevalent in only 17.4% when MSCT-based TAV sizing was performed (16.7% balloon-expandable THV vs. 17.6% self-expandable THV, p = 0.99). MSCT sizing was associated with reduced AR on multivariate analysis (odds ratio [OR]: 0.19, 95% confidence intervals [CI]: 0.08 to 0.45; p < 0.0001). Thirty-day device safety, success, and efficacy were noted in 79.1%, 89.9%, and 84.9% of patients, respectively. One-year mortality was 17.5%. Major vascular complications were associated with increased 1-year mortality (OR: 5.66, 95% CI: 1.21 to 26.43; p = 0.03).

Conclusions: TAV-in-BAV is feasible with encouraging short- and intermediate-term clinical outcomes. Importantly, a high incidence of post-implantation AR is observed, which appears to be mitigated by MSCT-based TAV sizing. Given the suboptimal echocardiographic results, further study is required to evaluate long-term efficacy.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jacc.2014.09.039DOI Listing

Publication Analysis

Top Keywords

aortic valve
16
self-expandable thv
16
type bav
12
msct-based tav
12
tav sizing
12
balloon-expandable thv
12
transcatheter aortic
8
bicuspid aortic
8
clinical outcomes
8
thv
8

Similar Publications

: New-onset postoperative atrial fibrillation (POAF) is the most common complication after cardiac surgery, occurring approximately in one-third of the patients. This study considered all-comer patients who underwent cardiac surgery to build a predictive model for POAF. : A total of 3467 (Center 1) consecutive patients were used as a derivation cohort to build the model.

View Article and Find Full Text PDF

Paravalvular leak (PVL) was initially recognized as one of the most common complications after transcatheter aortic valve implantation (TAVI) and has been linked to adverse clinical outcomes, including mortality. This study aims to assess the long-term clinical effects of PVL in patients undergoing TAVI with the latest generation of transcatheter aortic valves, as part of the national observational prospective multicenter study OBSERVANT II. OBSERVANT II included all consecutive patients with severe aortic stenosis who underwent TAVI across 28 Italian centers from December 2016 to September 2018.

View Article and Find Full Text PDF

Previous studies evaluated the prognostic role of hematological parameters in predicting outcome in patients with infective endocarditis (IE). However, only a few studies evaluated the role of hematological parameters in patients undergoing surgery for IE. The aim of this study was to review our 20-year experience with the treatment of native (NVE) and prosthetic (PVE) valve endocarditis and to evaluate the role of neutrophil-to-lymphocyte ratio (NLR), neutrophil-to-platelet ratio (NPR), platelet-to-lymphocyte ratio (PLR), monocyte-to-lymphocyte ratio (MLR), neutrophil-to monocyte ratio (NMR), and systemic inflammatory index (SII) on early and long-term outcomes of patients undergoing surgery for NVE and PVE.

View Article and Find Full Text PDF

Stuck prosthetic valves, often resulting from pannus formation or thrombus accumulation, represent a critical complication in prosthetic valve management, carrying significant risks for morbidity and mortality. This study aims to identify factors associated with stuck valve development and assess the effectiveness of interventions in restoring normal valve function. A total of 27 patients with stuck valves were analyzed, including mitral, aortic, and tricuspid valve cases.

View Article and Find Full Text PDF

Myocardial work (MW) is a new echocardiographic parameter used in the assessment of cardiac energy expenditure. The aim of the current study was to evaluate changes in left ventricular MW parameters in patients with severe aortic stenosis undergoing transcatheter aortic valve implantation (TAVI). One hundred and thirty five consecutive patients who underwent TAVI at one center were evaluated before and after the procedure using transthoracic echocardiography (TTE) to assess the following MW indices: global constructive work (GCW), global wasted work (GWW), global work index (GWI) and global work efficiency (GWE).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!