Sorting a large set of heavily used LiF:Mg,Ti thermoluminescent detectors into repeatable subsets of similar response.

Appl Radiat Isot

Department of Nuclear Engineering and Radiological Sciences, University of Michigan, 2355 Bonisteel Blvd., Ann Arbor, MI 48109-2104, USA; Department of Physics University of Azad Jammu and Kashmir Muzaffarbad, 13100 Azad Kashmir, Pakistan.

Published: January 2015

A set of 920 heavily used LiF:Mg,Ti thermoluminescent dosimeters (TLDs) was placed into a polymethyl methacrylate (PMMA) plate attached to a 40×40×15cm PMMA phantom and irradiated to 4.52mGy using a Cs source. This was repeated three times to determine the mean and standard deviation of each TLD׳s sensitivity. Reader drift was tracked over time with 10 control dosimeters. Two test sets of 100 TLDs were divided into subsets with sensitivities within ±1% of their subset means. All dosimeters were re-irradiated four times to test the TLDs׳ response repeatability and determine the sensitivity uniformity within the subsets. Coefficients of variation revealed that, within a given subset, the dosimeters responded within ±2.5% of their subset mean in all calibrations. The coefficient of variation in any of the 200 TLDs׳ calibrations was below 6% across the four calibrations. The work validates the approach of performing three calibrations to separate heavily used and aged TLDs with overall sensitivity variations of ±25% into subsets that reproducibly respond within ±2.5%.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.apradiso.2014.10.019DOI Listing

Publication Analysis

Top Keywords

heavily lifmgti
8
lifmgti thermoluminescent
8
subset dosimeters
8
sorting large
4
large set
4
set heavily
4
thermoluminescent detectors
4
detectors repeatable
4
subsets
4
repeatable subsets
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!