Enhanced quantification for 3D SEM-EDS: using the full set of available X-ray lines.

Ultramicroscopy

Department of Materials Science and Metallurgy, University of Cambridge, Charles Babbage Road 27, Cambridge, CB3 0FS Cambridgeshire, UK.

Published: January 2015

An enhanced method to quantify energy dispersive spectra recorded in 3D with a scanning electron microscope (3D SEM-EDS) has been previously demonstrated. This paper presents an extension of this method using all the available X-ray lines generated by the beam. The extended method benefits from using high energy lines, that are more accurately quantified, and from using soft X-rays that are highly absorbed and thus more surface sensitive. The data used to assess the method are acquired with a dual beam FIB/SEM investigating a multi-element Ni-based superalloy. A high accelerating voltage, needed to excite the highest energy X-ray line, results in two available X-ray lines for several elements. The method shows an improved compositional quantification as well as an improved spatial resolution.