Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Background: Accurate high-density lipoprotein cholesterol (HDL-C) measurements are important for management of cardiovascular diseases. The US Centers for Disease Control and Prevention (CDC) and Cholesterol Reference Method Laboratory Network (CRMLN) perform ultracentrifugation (UC) reference measurement procedure (RMP) to value assign HDL-C. Japanese CRMLN laboratory (Osaka) concurrently runs UC procedure and the designated comparison method (DCM). Osaka performance of UC and DCM was examined and compared with CDC RMP.
Methods: CDC RMP involved UC, heparin-MnCl₂ precipitation, and cholesterol analysis. CRMLN DCM for samples containing <200 mg/dl triglycerides involved 50-kDa dextran sulfate-MgCl2 precipitation and cholesterol determination.
Results: HDL-C regression equations obtained with CDC (x) and Osaka (y) were y=0.992x+0.542 (R(2)=0.996) for Osaka UC and y=1.004x-0.181 (R(2)=0.998) for DCM. Pass rates within ±1 mg/dl of the CDC target value were 91.9 and 92.1% for Osaka UC and DCM, respectively. Biases at 40 mg/dl HDL-C were +0.22 and -0.02 mg/dl for Osaka UC and DCM, respectively.
Conclusions: Osaka UC and DCM were highly accurate, precise, and stable for many years, assisting manufacturers to calibrate products for clinical laboratories to accurately measure HDL-C for patients, calculate non-HDL-C, and estimate low-density lipoprotein cholesterol with the Friedewald equation.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5695550 | PMC |
http://dx.doi.org/10.1016/j.cca.2014.10.039 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!