Cone-beam CT versus multi-slice CT systems for postoperative imaging of cochlear implantation--a phantom study on image quality and radiation exposure using human temporal bones.

Otol Neurotol

*Department of Otorhinolaryngology and Head and Neck Surgery, Radboud university medical center and Hearing and Implants, Donders Institute for Brain, Cognition and Behaviour, Nijmegen, The Netherlands; †Department of Radiology, Leiden University Medical Centre, Leiden, The Netherlands; ‡Department of Oral and Maxillofacial Surgery, 3D Facial Imaging Research Group, Radboud university medical center, Nijmegen, The Netherlands; and §Department of Radiology, Radboud university medical center, Nijmegen, The Netherlands.

Published: April 2015

Hypothesis: Image quality of low-dose multi-slice computed tomography (MSCT) after cochlear implantation is comparable to that of cone-beam computed tomography (CBCT).

Background: CBCT has been described as a low-dose alternative with superior image quality to MSCT for postoperative cochlear implant (CI) imaging, but to our knowledge, no dose-matched comparisons of image quality have been published.

Materials And Methods: Five human cochleae were implanted with CI electrodes and scanned on two CBCT and two MSCT systems. Four independent observers rated aspects of image quality on a five-point scale. CBCT scans were compared to clinical and dose-matched MSCT scans. Declining-dose MSCT protocols were compared to the clinical protocol. CT phantoms were used to determine effective dose and resolution for each acquisition protocol.

Results: Effective dose of the CBCT protocols was 6 to 16% of the clinical MSCT dose. Visibility of cochlear inner and outer walls and overall image quality were positively correlated with radiation dose on MSCT and image quality was better with clinical MSCT than with CBCT protocols. In other comparisons, differences between systems were found, but a distinction between CBCT and MSCT could not be made.

Conclusion: CBCT and dose-matched MSCT are both suitable for postoperative CI imaging. Selecting a CT system and radiation dose depends on which cochlear structures need to be visualized.

Download full-text PDF

Source
http://dx.doi.org/10.1097/MAO.0000000000000673DOI Listing

Publication Analysis

Top Keywords

image quality
28
msct
10
postoperative imaging
8
computed tomography
8
cbct msct
8
compared clinical
8
dose-matched msct
8
effective dose
8
cbct protocols
8
clinical msct
8

Similar Publications

The timely detection of viral pathogens in vineyards is a critical aspect of management. Diagnostic methods can be labor-intensive and may require specialized training or facilities. The emergence of artificial intelligence (AI) has the potential to provide innovative solutions for disease detection but requires a significant volume of high-quality data as input.

View Article and Find Full Text PDF

Pelvic Venous Disorder (PEVD) and May-Thurner syndrome (MTS) represent relatively understudied vascular issues that can significantly impact patients' quality of life. This study aims to evaluate the efficacy of surgical treatment for PEVD and MTS, conduct a comparative analysis of outcomes, and determine the practical significance of different therapeutic approaches. The study was conducted from 2019 to 2022 in Moscow, Russia, encompassing two outpatient clinics.

View Article and Find Full Text PDF

Nexus: A versatile console for advanced low-field MRI.

Magn Reson Med

January 2025

Department 8.1 - Biomedical Magnetic Resonance, Physikalisch-Technische Bundesanstalt (PTB), Braunschweig and Berlin, Germany.

Purpose: To develop a low-cost, high-performance, versatile, open-source console for low-field MRI applications that can integrate a multitude of different auxiliary sensors.

Methods: A new MR console was realized with four transmission and eight reception channels. The interface cards for signal transmission and reception are installed in PCI Express slots, allowing console integration in a commercial PC rack.

View Article and Find Full Text PDF

The subject of this paper is modeling customer satisfaction in the mobile telecommunication industry following the Covid-19 pandemic. Based on standard customer satisfaction models, a specialized model tailored for the mobile telecommunication industry has been developed to account for its unique characteristics, including market concentration. This model was created within the Slovakian context using the Structural Equation Modelling method.

View Article and Find Full Text PDF

Importance: In the Atrial Cardiopathy and Antithrombotic Drugs in Prevention After Cryptogenic Stroke (ARCADIA) randomized clinical trial, anticoagulation did not prevent recurrent stroke among patients with a recent cryptogenic stroke and atrial cardiopathy. It is unknown whether anticoagulation prevents covert infarcts in this population.

Objective: To test the use of apixaban vs aspirin for prevention of nonlacunar covert infarcts after cryptogenic stroke in patients with atrial cardiopathy.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!