Computer control of drug delivery by continuous intravenous infusion: bridging the gap between intended and actual drug delivery.

Anesthesiology

From the Division of Pulmonary, Critical Care, and Sleep Medicine, Department of Medicine, Beth Israel Deaconess Medical Center, Boston, Massachusetts (M.J.P.); Department of Anesthesiology and Pain Medicine, Steward St. Elizabeth's Medical Center, Boston, Massachusetts (M.A.L., A.E.W., M.G.W., M.Y.M.); Division of Pulmonology, Department of Medicine, Steward St. Elizabeth's Medical Center, Boston, Massachusetts (H.T.); and Department of Anesthesia, Critical Care, and Pain Medicine, Massachusetts General Hospital, Boston, Massachusetts (A.C.T., R.A.P.).

Published: March 2015

Background: Intravenous drug infusion driven by syringe pumps may lead to substantial temporal lags in achieving steady-state delivery at target levels when using very low flow rates ("microinfusion"). This study evaluated computer algorithms for reducing temporal lags via coordinated control of drug and carrier flows.

Methods: Novel computer control algorithms were developed based on mathematical models of fluid flow. Algorithm 1 controlled initiation of drug infusion and algorithm 2 controlled changes to ongoing steady-state infusions. These algorithms were tested in vitro and in vivo using typical high and low dead volume infusion system architectures. One syringe pump infused a carrier fluid and a second infused drug. Drug and carrier flowed together via a manifold through standard central venous catheters. Samples were collected in vitro for quantitative delivery analysis. Parameters including left ventricular max dP/dt were recorded in vivo.

Results: Regulation by algorithm 1 reduced delivery delay in vitro during infusion initiation by 69% (low dead volume) and 78% (high dead volume). Algorithmic control in vivo measuring % change in max dP/dt showed similar results (55% for low dead volume and 64% for high dead volume). Algorithm 2 yielded greater precision in matching the magnitude and timing of intended changes in vivo and in vitro.

Conclusions: Compared with conventional methods, algorithm-based computer control of carrier and drug flows can improve drug delivery by pump-driven intravenous infusion to better match intent. For norepinephrine infusions, the amount of drug reaching the bloodstream per time appears to be a dominant factor in the hemodynamic response to infusion.

Download full-text PDF

Source
http://dx.doi.org/10.1097/ALN.0000000000000519DOI Listing

Publication Analysis

Top Keywords

dead volume
20
computer control
12
drug delivery
12
low dead
12
drug
10
control drug
8
intravenous infusion
8
drug infusion
8
temporal lags
8
drug carrier
8

Similar Publications

Aims: To investigate the distribution of left atrioventricular coupling index (LACI) among patients with heart failure and left ventricular ejection fraction (LVEF)<50% and to explore its association with the combined endpoint of all-cause death or HF hospitalization at long term follow-up.

Methods And Results: Patients with HF and LVEF<50% undergoing cardiac magnetic resonance (CMR) were evaluated. Patients with atrial fibrillation or flutter were excluded.

View Article and Find Full Text PDF

There should be no assumption that an athlete is immune to coronary artery disease (CAD), even when traditional cardiovascular (CV) risk factors appear well-managed. Excelling in certain aspects of health does not equate to total CV protection. Recent data from cardiac imaging studies have raised the possibility that long-term, high-volume, high-intensity endurance exercise is associated with coronary atherosclerosis.

View Article and Find Full Text PDF

Objective: Patients with chronic kidney disease (CKD) have an increased risk of adverse cardio-cerebrovascular events. The purpose of this study is to evaluate the prognostic predictors over 5 years in patients with CKD including haemodialysis.

Methods: In this multicenter, prospective cohort study performed with the Gunma-CKD SPECT Study protocol, 311 patients with CKD [estimated glomerular filtration rate (eGFR) < 60 min/ml/1.

View Article and Find Full Text PDF

Edaravone Improves Motor Dysfunction Following Brachial Plexus Avulsion Injury in Rats.

ACS Chem Neurosci

January 2025

Department of Neurology, Multi-Omics Research Center for Brain Disorders,The First Affiliated Hospital, Hengyang Medical School, University of South China, Hengyang, Hunan 421001, China.

Brachial plexus root avulsion (BPRA) is often caused by road collisions, leading to total loss of motor function in the upper limb. At present, effective treatment options remain limited. Edaravone (EDA), a substance that eliminates free radicals, exhibits numerous biological properties, including neuroprotective, antioxidant and anti-inflammatory effects.

View Article and Find Full Text PDF

Chronic obstructive lung disease is the third leading cause of death worldwide. It affects the airways and lung parenchyma leading to emphysema. Bronchoscopic lung volume reduction is another strategy that aims to reduce air trapping and hyperinflation, leading to improvement in symptoms and pulmonary function.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!