Background: While spectral CT using single photon counting detector has shown a number of advantages in diagnostic imaging, knowledge of the detector response function of an energy-resolved detector is needed to correct the signal bias and reconstruct the image more accurately.

Objective: The objective of this paper is to study the photo counting detector response function using laboratory sources, and investigate the signal bias correction method.

Methods: Our approach is to model the detector response function over the entire diagnostic energy range (20 keV
Results: The analytical model generally describes the detector response function and is in good agreement with the measured data. The trend lines of the 12 parameters indicate higher energies tend to cause grater spectrum distortion. The spectrum distortion caused by the detector response function on spectral CT reconstruction is analyzed theoretically, and a solution to correct this spectrum distortion is also proposed.

Conclusion: In spectral and fluorescence CT, the spectrum distortion caused by detector response function poses a problem and cannot be ignored in any quantitative analysis. The detector response function of a CdTe detector can be obtained by a semi-analytical method.

Download full-text PDF

Source
http://dx.doi.org/10.3233/XST-140458DOI Listing

Publication Analysis

Top Keywords

detector response
16
response function
16
counting detector
12
function energy-resolved
8
single photon
8
photon counting
8
signal bias
8
detector
7
function
4
energy-resolved cdte
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!