Relationship between catheter contact force and radiofrequency lesion size and incidence of steam pop in the beating canine heart: electrogram amplitude, impedance, and electrode temperature are poor predictors of electrode-tissue contact force and lesion size.

Circ Arrhythm Electrophysiol

From the Heart Rhythm Institute (A.I., H.N., T.S., R.L., W.M.J.) and Department of Pathology, Veterans Administration Medical Center (J.V.P.), University of Oklahoma Health Sciences Center, Oklahoma City; St. Jude Medical GVA, Geneva, Switzerland (H.L., E.F., A.Y.); and University of Geneva, Geneva, Switzerland (D.C.S.).

Published: December 2014

Background: Electrode-tissue contact force (CF) is believed to be a major factor in radiofrequency lesion size. The purpose of this study was to determine, in the beating canine heart, the relationship between CF and radiofrequency lesion size and the accuracy of predicting CF and lesion size by measuring electrogram amplitude, impedance, and electrode temperature.

Methods And Results: Eight dogs were studied closed chest. Using a 7F catheter with a 3.5 mm irrigated electrode and CF sensor (TactiCath, St. Jude Medical), radiofrequency applications were delivered to 3 separate sites in the right ventricle (30 W, 60 seconds, 17 mL/min irrigation) and 3 sites in the left ventricle (40 W, 60 seconds, 30 mL/min irrigation) at (1) low CF (median 8 g); (2) moderate CF (median 21 g); and (3) high CF (median 60 g). Dogs were euthanized and lesion size was measured. At constant radiofrequency and time, lesion size increased significantly with increasing CF (P<0.01). The incidence of a steam pop increased with both increasing CF and higher power. Peak electrode temperature correlated poorly with lesion size. The decrease in impedance during the radiofrequency application correlated well with lesion size for lesions in the left ventricle but less well for lesions in the right ventricle. There was a poor relationship between CF and the amplitude of the bipolar or unipolar ventricular electrogram, unipolar injury current, and impedance.

Conclusions: Radiofrequencylesion size and the incidence of steam pop increase strikingly with increasing CF. Electrogram parameters and initial impedance are poor predictors of CF for radiofrequency ablation.

Download full-text PDF

Source
http://dx.doi.org/10.1161/CIRCEP.113.001094DOI Listing

Publication Analysis

Top Keywords

lesion size
28
contact force
12
radiofrequency lesion
12
beating canine
8
canine heart
8
electrogram amplitude
8
amplitude impedance
8
impedance electrode
8
electrode-tissue contact
8
ventricle seconds
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!