Objectives: To determine the implications of stable coverage of the coronary ostia by the Edwards SAPIEN valve frame in terms of myocardial ischemia and subsequent percutaneous coronary intervention (PCI), following transcatheter aortic valve implantation (TAVI).

Background: Edwards SAPIEN frame is frequently deployed relatively higher than recommended and may overlap the coronary ostia.

Methods: A total of 142 patients (age 81 ± 7 years, male 49%) treated with Edwards SAPIEN valve and with multi-detector row computed tomography at 1 month follow-up were evaluated. The position of the frame in relation to the coronary ostia was assessed. Levels of troponin T were measured 12-24 hr after TAVI. PCI events at follow-up were recorded.

Results: The left coronary ostium was fully covered in three (2.1%) patients and the right coronary ostium in 11 (7.7%). There were no differences in troponin T levels between patients with fully covered ostia versus patients with partly or non-covered ostia (0.24 (0.13-0.50) μg/L versus 0.35 (0.15-0.55) μg/L, respectively; P = 0.377). At 30 ± 15 months follow up, 10 (7%) patients underwent successful PCI. Rate of subsequent PCI was similar between patients with any covered ostium and patients with non-covered ostia [4 (7.8%) vs. 6 (6.5%), P = 0.780, respectively].

Conclusions: Full overlap of the coronary ostia by Edwards SAPIEN frame is infrequent and in most cases does not limit subsequent PCI.

Download full-text PDF

Source
http://dx.doi.org/10.1002/ccd.25718DOI Listing

Publication Analysis

Top Keywords

edwards sapien
20
coronary ostia
16
coronary
9
relation coronary
8
percutaneous coronary
8
ostia edwards
8
sapien valve
8
sapien frame
8
overlap coronary
8
coronary ostium
8

Similar Publications

Background: Transcatheter aortic valve implantation (TAVI) has shown similar or improved clinical outcomes compared with surgical aortic valve replacement (SAVR) in patients with symptomatic severe aortic stenosis at low risk for surgical mortality. This cost-utility analysis compared TAVI with SAPIEN 3 versus SAVR in symptomatic severe aortic stenosis patients at low risk of surgical mortality from the perspective of the Swedish healthcare system.

Methods: A published, two-stage, Markov-based cost-utility model that captured clinical outcomes from the (SWEDEHEART) registry (2018-2020) was adapted from the perspective of the Swedish healthcare system using local general population mortality, utility and costs data.

View Article and Find Full Text PDF

Background: The Edwards Sapien percutaneous valve (Edwards Lifesciences, Irvine, CA, USA) is a promising therapeutic option for congenital atrioventricular disease mostly because of the possibility to accommodate somatic growth with balloon dilatation.

Case Summary: This article reports the performance of the Edwards Sapien 3 valve in atrioventricular valve position in four paediatric patients.

Discussion: Despite aggressive antiplatelet and anticoagulation strategies, most patients showed early bioprosthesis dysfunction, with increasing gradient not related with somatic growth.

View Article and Find Full Text PDF

Background: Balloon-assisted bioprosthetic or native aortic scallop intentional laceration to prevent iatrogenic coronary artery obstruction (BA-BASILICA) enables valve-in-valve transcatheter aortic valve implantation (ViV-TAVI) in patients at risk of coronary artery obstruction. However, its efficacy in patients with severely calcified leaflets remains unclear.

Case Summary: We report a 78-year-old woman with a deteriorated 21 mm Carpentier-Edwards PERIMOUNT Magna valve.

View Article and Find Full Text PDF

Background: No data compare newer-generation transcatheter heart valves (THVs) in terms of next-day discharge (NDD) following transfemoral (TF) transcatheter aortic valve implantation (TAVI).

Aims: We aimed to evaluate the safety of NDD in unselected patients who received ACURATE (neo/neo2), Evolut (PRO/PRO+/FX) and the SAPIEN (3/Ultra) THVs.

Methods: This multicentre registry included patients who underwent TF-TAVI without a preprocedural permanent pacemaker implantation (PPI) and were discharged the next day without a new PPI.

View Article and Find Full Text PDF

Background: Transcatheter aortic valve replacement (TAVR)-in-TAVR within index supra-annular transcatheter heart valves (THVs) and high-risk anatomy can compromise coronary perfusion and re-access.

Case Summary: An 80-year-old male presented with acute heart failure caused by degeneration of an Evolut R THV, leading to severe aortic regurgitation. Aortic computed tomography imaging revealed a high-risk anatomy characterized by the index THV's commissural plane being placed above the coronary ostia and a valve-to-aorta distance of less than 2 mm, resulting not eligible for redo-TAVR according to recent literature.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!