Factors influencing accuracy and reproducibility of body resistance measurements by foot-to-foot impedancemeters.

J Med Eng Technol

UMR CNRS 7338, Department of Biological Engineering, Technological University of Compiegne, BP 20529, 60205 Compiegne , France.

Published: January 2015

The electronics of a BodySignal V2 (Tefal, France) foot-to-foot impedancemeter (FFI) was modified to display the foot-to-foot resistance instead of body fat. This device was connected to electrodes of different sizes mounted on a podoscope permitting photographs of subjects feet soles and electrodes in order to calculate the contact area between feet and electrodes. The foot-to-foot resistance was found to decrease when the contact area of feet with current and voltage electrodes increased. It was also sensitive to feet displacement and a backward move of 5 cm increased the mean resistance by 37 Ω. The resistance reproducibility was tested by asking the subject to repeat measurements 10-times by stepping up and down from the podoscope. The mean SD of these tests was 0.88% of mean resistance, but it fell to 0.47% when feet position was guided and to 0.29% with transverse voltage electrodes. For good reproducibility, it is important that voltage electrodes be small and that the scale design facilitates a correct position of heels on these electrodes.

Download full-text PDF

Source
http://dx.doi.org/10.3109/03091902.2014.973617DOI Listing

Publication Analysis

Top Keywords

voltage electrodes
12
foot-to-foot resistance
8
contact area
8
area feet
8
electrodes
7
resistance
6
feet
5
factors influencing
4
influencing accuracy
4
accuracy reproducibility
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!