Objective: To describe three DBS cases which presented with new side effects or loss of benefit from stimulation after long-term follow-up and to discuss the potential contributing factors.
Methods: A University of Florida (UF) database (INFORM) search was performed, identifying three patients, two Parkinson's disease (PD) and one Essential Tremor (ET), with an unexpected change in long-term programming thresholds as compared to initial evaluation. Clinical follow-up, programming, imaging studies, and lead measurements were reviewed. The UF Institutional Review Board (IRB) approved this study.
Results: A substantial increase in the 3rd ventricular width (120%), Evans index (6%), ventricular index (5%), and cella media index (17%) was uncovered. A change in thresholds across lead contacts with a decrease in current densities as well as a relative lateral change of lead location was also observed. Hardware-related complications, lead migration, and impedance variability were not identified.
Conclusions: Potential factors contributing to long-term side effects should be examined during a DBS troubleshooting assessment. Clinicians should be aware that in DBS therapy there is delivery of electricity to a changing brain, and atrophy may possibly affect DBS programming settings as part of long-term follow-up.