Undetected oesophageal perforation and feeding-tube misplacement.

Br J Nurs

Research Dietitian, Department of Nutrition and Dietetics, Southmead Hospital, Bristol.

Published: March 2015

This is a case report of an electromagnetically (EM)-guided Cortrak feeding tube that perforated the lower oesophagus and was not detected by the EM trace or by plain X-ray. Misplacement was diagnosed from computed tomography (CT) following injection of radio-contrast medium down the tube. Recommendations are offered for use of the EM trace in patients at high risk of oesophageal perforation.

Download full-text PDF

Source
http://dx.doi.org/10.12968/bjon.2014.23.19.1020DOI Listing