Objective: Laser surgery of the larynx and airway remains high risk for the formation of operating room fire. Traditional methods of fire prevention have included use of "laser safe" tubes, inflation of a protective cuff with saline, and wet pledgets to protect the endotracheal tube from laser strikes. We tested a mechanical model of laser laryngeal surgery to evaluate the fire risk.
Study Design: Mechanical model.
Setting: Laboratory.
Subjects And Methods: An intubation mannequin was positioned for suspension microlaryngoscopy. A Laser-Shield II cuffed endotracheal tube was placed through the larynx and the cuff inflated using saline. Wet pledgets covered the inflated cuff. A CO2 laser created an inadvertent cuff strike at varying oxygen concentrations. Risk reduction measures were implemented to discern any notable change in the outcome after fire.
Results: At 100% FiO2 an immediate fire with sustained flame was created and at 40% FiO2 a near immediate sustained flame was created. At 29% FiO2, a small nonsustained flame was noted. At room air, no fire was created. There was no discernible difference in the severity of laryngeal damage after the fire occurred whether the tube was immediately pulled from the mannequin or if saline was poured down the airway as a first response.
Conclusions: While "laser safe" tubes provide a layer of protection against fires, they are not fire proof. Inadvertent cuff perforation may result in fire formation in low-level oxygen enriched environments. Placement of wet pledgets do not provide absolute protection. Endotracheal tube (ETT) cuffs should be placed distally well away from an inadvertent laser strike while maintaining the minimum supplemental oxygen necessary.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1177/0194599814555853 | DOI Listing |
Ann Otol Rhinol Laryngol
May 2021
Department of Otolaryngology, Henry Ford Health System, Detroit, MI, USA.
Background: Intranasal topical 1:1000 epinephrine has been used safely and effectively for hemostasis during endoscopic sinus surgery (ESS). Prior studies assessing hemodynamic changes after intranasal topical epinephrine application have only used soaking wet cottonoid pledgets, and have only assessed for hemodynamic changes before any surgery being performed.
Objective: The purposes of this study were to determine whether intranasal application of topical 1:1000 epinephrine with wrung-out cottonoid pledgets caused significant hemodynamic changes both before and during ESS, and whether it allowed for adequate hemostasis.
Otolaryngol Head Neck Surg
January 2015
Hofstra North Shore-LIJ School of Medicine, Chief of Pediatric Otolaryngology, Cohen Children's Medical Center, Long Island, New York, USA.
Objective: Laser surgery of the larynx and airway remains high risk for the formation of operating room fire. Traditional methods of fire prevention have included use of "laser safe" tubes, inflation of a protective cuff with saline, and wet pledgets to protect the endotracheal tube from laser strikes. We tested a mechanical model of laser laryngeal surgery to evaluate the fire risk.
View Article and Find Full Text PDFJ Neurol Surg A Cent Eur Neurosurg
November 2013
Department of Neurosurgery, Yamagata University Faculty of Medicine, Iidanishi, Yamagata, Japan.
Background: Intraoperative high-field magnetic resonance imaging (iMRI) is a useful modality for immediate intraoperative quality control. With iMRI, a surgeon can confirm whether tumor remnants exist during surgery; which makes it possible to add further resection, obtain a higher resection rate, and improve the cure rate. It is sometimes difficult to evaluate the existence of tumor remnants when the tumor resection cavity is collapsed.
View Article and Find Full Text PDFNeurosurgery
February 2006
Department of Neurosurgery, Mount Sinai School of Medicine, Hollis Hills, New York, USA.
A major concern in surgery is the prevention or control of bleeding. The ligature and the clip are the hallmarks of the last century of modern surgery. Therapeutic embolization is not really used to stop hemorrhage but to treat aneurysms and so prevent future rupture and bleeding.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!