Intraocular lens opacification after descemet stripping automated endothelial keratoplasty.

Cornea

Department of Cornea and External Eye Diseases, Bristol Eye Hospital, Bristol, United Kingdom.

Published: December 2014

Purpose: Intraocular lens (IOL) opacification is a rare but serious complication that may necessitate its exchange. The use of intraocular gases is a known precipitant. Descemet stripping endothelial keratoplasty (DSAEK) involves injecting air into the anterior chamber. IOL opacification has been described after this procedure; however its incidence is currently unknown.

Methods: A retrospective review of case notes from a single center of all patients undergoing DSAEK, who were either already pseudophakic or had simultaneous cataract surgery. Cases with IOL opacification were analyzed, and any risk factors were identified.

Results: One hundred sixty-eight DSAEK were performed on 154 eyes of 137 patients. Fifty-four cases had simultaneous cataract surgery with implantation of an IOL. Fifteen (9.7%) eyes developed IOL opacification. This had a distinctive pattern, being limited to the anterior lens surface, in the pupillary zone. Median time interval from keratoplasty to the first observation of IOL opacification was 17 months (range, 4-34 months). The only statistically significant risk factor was rebubbling of detached endothelial grafts. Rebubbling was performed in 62.5% (10/15) of cases with IOL opacification, compared with 23% (32/139) with no opacification (P = 0.0009).

Conclusions: This is the first study to report the incidence of IOL opacification after undergoing DSAEK. Although the mechanism is unknown, multiple injections of air into the anterior chamber statistically increase the risk of IOL opacification.

Download full-text PDF

Source
http://dx.doi.org/10.1097/ICO.0000000000000279DOI Listing

Publication Analysis

Top Keywords

iol opacification
32
opacification
10
iol
9
intraocular lens
8
descemet stripping
8
endothelial keratoplasty
8
air anterior
8
anterior chamber
8
undergoing dsaek
8
simultaneous cataract
8

Similar Publications

Purpose: Severely myopic eyes have been associated with high posterior capsule opacification (PCO) incidence. Although it has been reported that myopic eyes have weaker or more delayed capsule adhesion than emmetropic eyes, it is unclear whether/how dioptric power and posterior curvature of IOLs affect IOLs' affinity for the posterior lens capsule (PLC) and their PCO potential.

Methods: To investigate this, acrylic foldable IOLs with increasing dioptric power of 6.

View Article and Find Full Text PDF

Background: Pseudophakic cystoid macular edema (CME) following primary anterior-chamber intraocular lens (ACIOL) implantations is commonly seen. Intravitreal triamcinolone acetonide (IVTA) injections have shown significant improvement in visual acuity and retinal thickness in refractory pseudophakic CME. Pseudohypopyon following IVTA injection is a known entity.

View Article and Find Full Text PDF

This study aims to analyze the prevalence and severity of posterior capsule opacification (PCO) and glistening in a new hydrophobic biaspheric monofocal intraocular lens (IOLs) 24 months after implantation. By means of a ambispective, observational, case-control design, a total of 297 eyes from 200 cataract surgery patients were included in the study (118 females and 82 males; mean age: 72.31 ± 9.

View Article and Find Full Text PDF

Innovative Polymeric Biomaterials for Intraocular Lenses in Cataract Surgery.

J Funct Biomater

December 2024

Department of Surgery, Division of Ophthalmology, University of Sherbrooke, Sherbrooke, QC J1G 2E8, Canada.

Intraocular lenses (IOLs) play a pivotal role in restoring vision following cataract surgery. The evolution of polymeric biomaterials has been central to addressing challenges such as biocompatibility, optical clarity, mechanical stability, and resistance to opacification. This review explores essential requirements for IOL biomaterials, emphasizing their ability to mitigate complications like posterior capsule opacification (PCO) and dysphotopsias while maintaining long-term durability and visual quality.

View Article and Find Full Text PDF

Scleral fixated intraocular lens techniques for the vitreoretinal surgeon: an update.

Curr Opin Ophthalmol

December 2024

Department of Research, Wills Eye Hospital, Philadelphia, PA.

Purpose Of Review: To describe recent evidence regarding surgical techniques for secondary scleral fixation of intraocular lenses (IOLs), with attention to information most relevant to the surgical practice of vitreoretinal surgeons.

Recent Findings: Recent studies have reported favorable outcomes in patients treated with scleral fixation of IOLs using Gore-Tex suture and flanged intrascleral haptic fixation (ISHF). ISHF can be achieved using trocar-cannulas or double needle-based techniques.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!