A PHP Error was encountered

Severity: Warning

Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests

Filename: helpers/my_audit_helper.php

Line Number: 176

Backtrace:

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML

File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global

File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword

File: /var/www/html/index.php
Line: 316
Function: require_once

Spinal cord stimulation for the treatment of chronic back pain patients: 500-Hz vs. 1000-Hz burst stimulation. | LitMetric

Objective: Spinal cord stimulation is a commonly used, safe, and effective procedure applied for medically intractable failed back surgery syndrome, as well as other neuropathic pain syndromes. Recently, a novel stimulation paradigm called burst stimulation has been developed that is paresthesia-free and has a more pronounced suppressive effect on neuropathic pain.

Materials And Methods: Fifteen patients who were being treated with burst spinal cord stimulation for failed back surgery syndrome participated in an open-label trial to verify whether their pain suppression could be further ameliorated by changing the burst pattern. Burst stimulation with packets of five electrical pulses delivered at 500 Hz with 1000-μsec pulse width 40 times per second was changed to burst mode delivering five spikes at 1000 Hz with 500-μsec pulse width 40 times a second. As the amplitudes did not differ between the two groups, the total delivery of current to the spinal cord was not different between the two modes of burst stimulation. Scores on visual analog scales for pain and paresthesia, the Pain Catastrophizing Scale, the Pain Vigilance and Awareness Questionnaire, and the Short Form 36 quality of life measurement were compared between the two modes of burst stimulation. [Correction added on 06 Feb 2015, after first online publication: this paragraph has been revised to signify the comparison of amplitudes between two groups]

Results: No statistically significant differences were found between the two modes of stimulation.

Conclusion: The results suggest that increasing the frequency from 500 to 1000 Hz while keeping the pulse width constant does not add any extra benefit in suppressing pain. Further studies should verify whether increasing the frequency above 1000 Hz has a similar lack of effect.

Download full-text PDF

Source
http://dx.doi.org/10.1111/ner.12252DOI Listing

Publication Analysis

Top Keywords

burst stimulation
20
spinal cord
16
cord stimulation
12
pulse width
12
stimulation
9
burst
8
failed surgery
8
surgery syndrome
8
width times
8
times second
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!