Introduction: Early ceramic bearing systems in total hip arthoplasty (THA) sought to provide long term wear improvement over traditional metal on polyethylene systems. However, previous designs exhibited fractures of the ceramic acetabular liner, leading to the development of the Implex Hedrocel ceramic bearing THA system where the ceramic liner was supported on a layer of polyethylene intended to transition liner loads to the metal shell, a so-called "sandwich" design. Unfortunately, the device trial was stopped to further enrollment when liner fractures were reported. The current study examines nearly 10-year follow-up on 28 devices implanted by two surgeons at one institution in order to document ceramic bearing system performance over a longer time period.

Methods: Radiographic and patient reported outcomes, in the form of Harris Hip Scores (HHS) and 12-Item Short Form Health Survey (sF-12), were collected.

Results: During the study period two cups were replaced, one at three years and a second at seven years. At the five year follow-up HHS were similar to those reported in the literature for devices with traditional metal-on-polyethylene bearing surfaces and for other sandwich ceramic bearing designs. At the nine year follow-up, the HHS had not changed significantly and SF-12 scores measuring overall physical and mental health were higher than age matched national norms (p<0.001). There were no signs of cup migration, stem subsidence, osteolysis or cup loosening at any time up to the last follow-up in this patient cohort. The 89% survivorship rate and device revisions due to delamination of the liner observed in this group were similar to those reported earlier for this device and for other "sandwich design" ceramic bearing systems.

Discussion: This cohort did not exhibit new failure modes and HHS and SF-12 scores indicated high functionality for the majority of patients. These data suggest that a focus on preventing ceramic liner fracture through design and/or materials improvements may result in a device with long-term functionality.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4127741PMC

Publication Analysis

Top Keywords

ceramic bearing
16
year follow-up
12
total hip
8
follow-up hhs
8
bearing
6
ceramic
6
follow-up ceramic-on-ceramic
4
ceramic-on-ceramic bearing
4
bearing total
4
hip arthroplasty
4

Similar Publications

The Guangyuan kiln, located in the Sichuan Province, Southwest China during the Song Dynasty (960-1279 A.D.), is renowned for its high-temperature iron-series glazed wares, including pure black glazed ware, hare's fur glazed ware, glossy brown glazed ware, and matte brown glazed ware.

View Article and Find Full Text PDF
Article Synopsis
  • The study evaluates the fracture strength of 1 mm-thick CAD/CAM occlusal veneers made from lithium disilicate (LD) and resin nanoceramics (RNC) to assess their viability in dental restoration.
  • Using a novel testing protocol, the RNC group demonstrated a significantly higher load-bearing capacity compared to the LD group, indicating better durability under stress.
  • Both materials experienced similar crack patterns during testing, emphasizing the importance of understanding mechanical properties for ensuring the longevity of dental restorations.
View Article and Find Full Text PDF

Atraumatic Fracture of Newer Generation Ceramic Head Three Days Post-op: A Case Report.

Cureus

December 2024

Orthopedic Surgery, Tri-County Orthopedics, Bridgewater, USA.

A 73-year-old female experienced an atraumatic fracture of a BIOLOX delta ceramic femoral head following uncomplicated right total hip arthroplasty using a ceramic-on-polyethylene bearing. The fracture occurred post-operatively, as revealed by radiography after the patient reported a clunking sensation and leg shortening. Revision surgery involved replacing the fractured head and liner with careful removal of ceramic debris.

View Article and Find Full Text PDF

This work aims to address key issues in the ballistic performance of ceramic-based composite armor, particularly at the joints of spliced ​​bulletproof panels. The edge structure of C/C-SiC ceramic plates and ultra-high molecular weight polyethylene is redesigned to superimpose the joint areas. These structurally optimized composite pads are examined by numerical simulation of impact dynamics to understand their anti-penetration performance whose accuracy is then validated by live fire tests.

View Article and Find Full Text PDF

Adverse local tissue reaction (ALTR) and osteolysis formation are more frequently described complications after total hip arthroplasty (THA) with metal-on-metal bearings. In this report, we present a unique case of ALTR and osteolysis formation with ceramic-on-ceramic bearing THA, which has been less commonly described in the existing literature. This rare case occurred in a 54-year-old patient who had a primary THA done 17 years prior to representing to our institution.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!