Objectives: Our two objectives were to evaluate the feasibility of fetal brain magnetic resonance imaging (MRI) using a fast spin echo sequence at 3.0T field strength with low radio frequency (rf) energy deposition (as measured by specific absorption rate: SAR) and to compare image quality, tissue contrast and conspicuity between 1.5T and 3.0T MRI.

Methods: T2 weighted images of the fetal brain at 1.5T were compared to similar data obtained in the same fetus using a modified sequence at 3.0T. Quantitative whole-body SAR and normalized image signal to noise ratio (SNR), a nominal scoring scheme based evaluation of diagnostic image quality, and tissue contrast and conspicuity for specific anatomical structures in the brain were compared between 1.5T and 3.0T.

Results: Twelve pregnant women underwent both 1.5T and 3.0T MRI examinations. The image SNR was significantly higher (P=0.03) and whole-body SAR was significantly lower (P<0.0001) for images obtained at 3.0T compared to 1.5T. All cases at both field strengths were scored as having diagnostic image quality. Images from 3.0T MRI (compared to 1.5T) were equal (57%; 21/37) or superior (35%; 13/37) for tissue contrast and equal (61%; 20/33) or superior (33%, 11/33) for conspicuity.

Conclusions: It is possible to obtain fetal brain images with higher resolution and better SNR at 3.0T with simultaneous reduction in SAR compared to 1.5T. Images of the fetal brain obtained at 3.0T demonstrated superior tissue contrast and conspicuity compared to 1.5T.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5987203PMC
http://dx.doi.org/10.1515/jpm-2014-0268DOI Listing

Publication Analysis

Top Keywords

fetal brain
12
15t 30t
12
image quality
12
brain 15t
8
30t field
8
specific absorption
8
absorption rate
8
rate sar
8
sequence 30t
8
quality tissue
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!