The effect of suture-button fixation on simulated syndesmotic malreduction: a cadaveric study.

J Bone Joint Surg Am

Department of Orthopaedic Surgery and Rehabilitation, University of Iowa Hospitals and Clinics, 200 Hawkins Drive, 01008 JPP, Iowa City, IA 52242. E-mail address for R.W. Westermann: E-mail address for C. Rungprai: E-mail address for J. Femino: E-mail address for A. Amendola: E-mail address for P. Phisitkul: E-mail address:

Published: October 2014

Background: The accuracy of reduction of distal tibiofibular syndesmosis disruptions has been associated with the clinical outcome. Suture-button fixation of the syndesmosis is a dynamic alternative mode of fixation. We hypothesized that with deliberate clamp-induced malreduction, suture-button fixation of the syndesmosis would allow a more anatomic post-fixation position compared with screw fixation.

Methods: Forty-eight syndesmotic fixations were performed on twelve through-knee cadaveric specimens. The syndesmosis was destabilized and off-axis clamping was used to produce both anterior and posterior malreduction patterns. In twelve scenarios (six anterior and six posterior malreductions), syndesmotic screw fixation was used, followed by computed tomography. With tenacula holding the malreduction, the syndesmosis screws were exchanged for a suture-button construct and the specimens underwent a subsequent computed tomography scan. In the other twelve scenarios, the suture-button fixation was achieved first, followed by screw fixation. Standardized measurements of anterior-posterior and medial-lateral fibular displacement were performed by two observers blinded to the method of fixation.

Results: With anterior off-axis clamping, the mean sagittal malreduction was 2.7 ± 2.0 mm with screw fixation and 1.0 ± 1.0 mm with suture-button fixation (p = 0.02). With posterior off-axis clamping, the sagittal malreduction was 7.2 ± 2.3 mm with screw fixation and 0.5 ± 1.4 mm with suture-button fixation (p < 0.01). No differences were observed between fixation types in the coronal plane (p = 0.20 for anterior malreductions and p = 0.06 for posterior malreductions).

Conclusions: With deliberate malreduction in a cadaver model, suture-button fixation of the syndesmosis results in less post-fixation displacement compared with screw fixation. The suture button's ability to allow for natural correction of deliberate malreduction was greatest with posterior off-axis clamping.

Clinical Relevance: Although the clinical relevance is unknown, dynamic syndesmotic fixation may mitigate clamp-induced malreduction.

Download full-text PDF

Source
http://dx.doi.org/10.2106/JBJS.N.00198DOI Listing

Publication Analysis

Top Keywords

suture-button fixation
28
screw fixation
20
fixation
14
fixation syndesmosis
12
off-axis clamping
12
malreduction
9
suture-button
8
clamp-induced malreduction
8
compared screw
8
anterior posterior
8

Similar Publications

Purpose: This study aims to describe a fixation technique for coronoid fractures using suture buttons, and to biomechanically evaluate this technique in comparison to screw fixation as a time-zero pilot study.

Methods: An O'Driscoll type 2 anteromedial coronoid facet (AMCF) fracture was simulated in 20 fresh-frozen human elbows. The specimens were randomized into two groups and fracture fixation was performed with either a suture button system or a 3.

View Article and Find Full Text PDF

Background: The Latarjet and other bony augmentation procedures are commonly used to treat anterior shoulder instability in the setting of significant glenoid bone loss. Although several fixation strategies have been reported, the biomechanical strength of these techniques remains poorly understood.

Purpose: To perform a systematic review of the biomechanical strength of glenoid bony augmentation procedures for anterior shoulder instability.

View Article and Find Full Text PDF

Background: This study aimed to describe the arthroscopic superlateral capsule pathway technique for spotting femoral fixation device deployment, and to compare the results with normal procedure.

Methods: A total of 69 patients underwent ACLR (Anterior Cruciate Ligament Reconstruction) with or without the SCP (superolateral capsule pathway) during procedure were retrospectively selected and evaluated. A total of 36 patients underwent SCP and 33 patients underwent ACLR without SCP.

View Article and Find Full Text PDF

Background: Iatrogenic suprascapular nerve injury secondary to posterior drilling or screw penetration is a recognized complication of bone block or coracoid process transfers for anterior glenohumeral instability. We present the first cadaveric study that assesses the safety of posteroanterior reference guides and quantifies the relationship of the suprascapular nerve to posterior glenoid fixation with suture buttons.

Methods: Anterior glenoid bone block reconstruction with suture buttons utilizing a posteroanterior reference guide was performed in 10 fresh frozen cadavers via a posterior portal.

View Article and Find Full Text PDF

» Suture button fixation has emerged as an effective surgical construct in arthroscopic and open Latarjet and anterior glenoid reconstruction with free autograft bone with high rates of bone block healing.» Biomechanical data suggest that screw and suture button mechanical fixation constructs provide similar load to failure and stability for the Latarjet procedure.» Preliminary bone healing models have identified that flexible fixation may exhibit a higher degree of bone callus maturation, whereas rigid fixation results in excessive callus hyperplasia» Mechanical tensioner use provides consistent tensioning of suture button constructs and improves bone-to-bone healing rates when used for anterior glenoid reconstruction surgery.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!