Objectives: Pitch plasticity has been observed in Hybrid cochlear implant (CI) users. Does pitch plasticity also occur in bimodal CI users with traditional long-electrode CIs, and is pitch adaptation pattern associated with electrode discrimination or speech recognition performance? The goals of this study were to characterize pitch adaptation patterns in long-electrode CI users, to correlate these patterns with electrode discrimination and speech perception outcomes, and to analyze which subject factors are associated with the different patterns.

Design: Electric-to-acoustic pitch matches were obtained in 19 subjects over time from CI activation to at least 12 months after activation, and in a separate group of 18 subjects in a single visit after at least 24 months of CI experience. Audiometric thresholds, electrode discrimination performance, and speech perception scores were also measured.

Results: Subjects measured over time had pitch adaptation patterns that fit one of the following categories: (1) "Pitch-adapting," that is, the mismatch between perceived electrode pitch and the corresponding frequency-to-electrode allocations decreased; (2) "Pitch-dropping," that is, the pitches of multiple electrodes dropped and converged to a similar low-pitch; and (3) "Pitch-unchanging," that is, the electrode pitches did not change. Subjects measured after CI experience had a parallel set of adaptation patterns: (1) "Matched-pitch," that is, the electrode pitch was matched to the frequency allocation; (2) "Low-pitch," that is, the pitches of multiple electrodes were all around the lowest frequency allocation; and (3) "Nonmatched-pitch," that is, the pitch patterns were compressed relative to the frequency allocations and did not fit either the matched-pitch or low-pitch categories. Unlike Hybrid CI users which were mostly in the pitch-adapting or matched-pitch category, the majority of bimodal CI users were in the latter two categories, pitch-dropping/low-pitch or pitch-unchanging/nonmatched-pitch. Subjects with pitch-adapting or matched-pitch patterns tended to have better low-frequency thresholds than subjects in the latter categories. Changes in electrode discrimination over time were not associated with changes in pitch differences between electrodes. Reductions in speech perception scores over time showed a weak but nonsignificant association with dropping-pitch patterns.

Conclusions: Bimodal CI users with more residual hearing may have somewhat greater similarity to Hybrid CI users and be more likely to adapt pitch perception to reduce mismatch with the frequencies allocated to the electrodes and the acoustic hearing. In contrast, bimodal CI users with less residual hearing exhibit either no adaptation, or surprisingly, a third pattern in which the pitches of the basal electrodes drop to match the frequency range allocated to the most apical electrode. The lack of association of electrode discrimination changes with pitch changes suggests that electrode discrimination does not depend on perceived pitch differences between electrodes, but rather on some other characteristics such as timbre. In contrast, speech perception may depend more on pitch perception and the ability to distinguish pitch between electrodes, especially since during multielectrode stimulation, cues such as timbre may be less useful for discrimination.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4336615PMC
http://dx.doi.org/10.1097/AUD.0000000000000114DOI Listing

Publication Analysis

Top Keywords

electrode discrimination
24
pitch
16
pitch adaptation
16
adaptation patterns
16
bimodal users
16
speech perception
16
electrode
10
users
9
cochlear implant
8
implant users
8

Similar Publications

Introduction: Conductive disturbances requiring permanent pacemaker (PPM) implantation remain a major concern after transcatheter aortic valve implantation (TAVI).

Aims: To assess the impact of aortic valve calcium score (AVCS) on conductive disturbances requiring PPM after TAVI.

Methods: All patients who underwent TAVI with accessible AVCS from the preprocedural CT scan report were included in this retrospective single-centre study.

View Article and Find Full Text PDF

Cell-free transcription amplification-based split-type electrochemical sensor using enzyme-linked magnetic microbeads for minimal residual leukemia detection.

Talanta

January 2025

Department of Rehabilitation Medicine, School of Health, Fujian Medical University, Fuzhou, 350122, China; Department of Orthopaedics, Fujian Provincial Institute of Orthopaedics, The First Affiliated Hospital, Fujian Medical University, Fuzhou, 350005, China. Electronic address:

Constrained by detecting techniques, patients with acute promyelocytic leukemia (APL) are often confronted with minimal residual disease (MRD) and a high risk of relapse. Thus, a pragmatic and robust method for MRD monitoring is urgently needed. Herein, a novel split-type electrochemical sensor (E-sensor) was developed by integrating nucleic acid sequence-based amplification (NASBA) with enzyme-linked magnetic microbeads (MMBs) for ultra-sensitive detection of the PML/RARα transcript.

View Article and Find Full Text PDF

Cochlear Implant Electrode Placement and Music Perception.

JAMA Otolaryngol Head Neck Surg

January 2025

Department of Hearing and Speech Sciences, Vanderbilt University Medical Center, Nashville, Tennessee.

Importance: Cochlear implants enable improvements in speech perception, but music perception outcomes remain variable. Image-guided cochlear implant programming has emerged as a potential programming strategy for increasing the quality of spectral information delivered through the cochlear implant to improve outcomes.

Objectives: To perform 2 experiments, the first of which modeled the variance in music perception scores as a function of electrode positioning factors, and the second of which evaluated image-guided cochlear implant programming as a strategy to improve music perception with a cochlear implant.

View Article and Find Full Text PDF

This paper presents a lens-free imaging approach utilizing an array of light sources, capable of measuring the dielectric properties of many particles simultaneously. This method employs coplanar electrodes to induce velocity changes in flowing particles through dielectrophoretic forces, allowing the inference of individual particle properties from differential velocity changes. Both positive and negative forces are detectable.

View Article and Find Full Text PDF

Early detection of hepatitis C virus (HCV) infection is crucial for eliminating this silent killer, especially in resource-limited settings. HCV core antigen (HCVcAg) represents a promising alternative to the current "gold standard" HCV RNA assays as an active viremia biomarker. Herein, a highly sensitive electrochemical magneto-immunosensor for the HCVcAg was developed.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!