Objectives: Our aim was to develop an aortic stent graft phantom to simulate endoleak treatment and to find a tantalum content (TC) of ethylene-vinyl-alcohol-copolymer that causes fewer computed tomography (CT) beam hardening artefacts, but still allows for fluoroscopic visualization.

Methods: Ethylene-vinyl-alcohol-copolymer specimens of different TC (10-50 %, and 100 %) were injected in an aortic phantom bearing a stent graft and endoleak cavities with simulated re-perfusion. Fluoroscopic visibility of the ethylene-vinyl-alcohol-copolymer specimens was analyzed. In addition, six radiologists analyzed endoleak visibility, and artefact intensity of ethylene-vinyl-alcohol-copolymer in CT.

Results: Reduction of TC significantly decreased CT artefact intensity of ethylene-vinyl-alcohol-copolymer and increased visibility of endoleak re-perfusion (p < 0.000). It also significantly decreased fluoroscopic visibility of ethylene-vinyl-alcohol-copolymer (R = 0.883, p ≤ 0.01), and increased the active embolic volumes prior to visualization (Δ ≥ 40 μl). Ethylene-vinyl-alcohol-copolymer specimens with a TC of 45-50 % exhibited reasonable visibility, a low active embolic volume and a tolerable CT artefact intensity.

Conclusions: The developed aortic stent graft phantom allows for a reproducible simulation of embolization of endoleaks. The data suggest a reduction of the TC of ethylene-vinyl-alcohol-copolymer to 45 -50 % of the original, to interfere less with diagnostic imaging in follow-up CT examinations, while still allowing for fluoroscopic visualization.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00330-014-3438-9DOI Listing

Publication Analysis

Top Keywords

beam hardening
8
hardening artefacts
8
tantalum content
8
stent graft
8
ethylene-vinyl-alcohol-copolymer specimens
8
artefact intensity
8
intensity ethylene-vinyl-alcohol-copolymer
8
endoleak
5
ethylene-vinyl-alcohol-copolymer
5
reduction beam
4

Similar Publications

Assessment of Corrosion and Cavitation Resistance of Laser Remelted GX40CrNiSi25-20 Cast Stainless Steel.

Materials (Basel)

December 2024

Department of Materials and Fabrication Engineering, Politehnica University Timisoara, Bulevardul Mihai Viteazul nr.1, 300222 Timișoara, Romania.

This paper explores the enhancement of cavitation and corrosion resistance in cast stainless steel through laser beam surface remelting. The influence of laser treatment on material properties was assessed by analyzing the microstructure using optical microscopy, electron microscopy, and X-ray diffraction. Cavitation erosion was evaluated in tap water using an ultrasonic vibration setup, following ASTM G32-2016 standards.

View Article and Find Full Text PDF

This study aimed to optimize CT esophagography by identifying effective oral contrast dilution ratios and exploring the advantages of dual-energy CT (DECT) over conventional CT for improving image quality. Ex vivo experiments using iodine contrast agents (320-400 mgI/mL) at 21 dilution ratios were scanned at three voltages, with additional dual-energy scans generating various reconstruction images. Image quality was assessed both objectively and subjectively.

View Article and Find Full Text PDF

Expression of beam hardening artifacts on horizontally stitched cone-beam computed tomography images.

Imaging Sci Dent

December 2024

Department of Oral Diagnosis, Division of Oral Radiology, Piracicaba Dental School, University of Campinas, Piracicaba, SP, Brazil.

Purpose: This study was performed to evaluate the expression of beam hardening artifacts generated by high atomic number materials in stitched cone-beam computed tomography (CBCT) images, compared to the traditional acquisition mode.

Materials And Methods: CBCT volumes were acquired using an acrylic resin phantom embedded with pairs of cylinders made from amalgam dental alloy, cobalt-chromium alloy, gutta-percha, titanium, and zirconium. These cylinders were placed within the overlapping zones of the stitching reconstruction area.

View Article and Find Full Text PDF

Purpose: Photon counting detectors offer promising advancements in computed tomography (CT) imaging by enabling the quantification and three-dimensional imaging of contrast agents and tissue types through simultaneous multi-energy projections from broad X-ray spectra. However, the accuracy of these decomposition methods hinges on precise composite spectral attenuation values that one must reconstruct from spectral micro-CT. Errors in such estimations could be due to effects such as beam hardening, object scatter, or detector sensor-related spectral distortions such as fluorescence.

View Article and Find Full Text PDF

Background: Percutaneous image-guided cementoplasty is a medical procedure for strengthening bones structurally altered by disease, such as osteolytic metastasis. This procedure involves injecting biocompatible liquid bone cement, through one or more trocars into the damaged bone. Within a few minutes the bone cement hardens and restores the rigidity of the bony structure.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!