Purpose: To assess the fatigue and in vivo performance of a new stent-graft incorporating bovine peritoneum lining that is designed for application in peripheral vascular occlusive disease.

Methods: Bovine peritoneum-lined stent-grafts were subjected to accelerated in vitro pulsatile fatigue and axial/torsional fatigue testing designed to simulate 10 years of physiological strain on the devices. At specified times the devices were evaluated for stent fracture, suture failure, or tissue tearing. Seven dogs underwent bilateral common iliac artery (CIA) balloon angioplasty injury with unilateral placement of the peritoneum-lined stent-graft. Angiography and intravascular ultrasound were performed prior to treatment, after treatment, and prior to sacrifice at 30 days. Vessels were perfusion fixed and histologically evaluated at 5 regions: above stent, proximal stent, mid stent, distal stent, and below stent.

Results: No evidence of stent, suture, or tissue failure was present during or after pulsatile and axial/torsional fatigue testing. At 30±0.3 days after implantation, all vessels were patent. The average lumen area at explantation across stented vessels was 25.45 mm(2). Lumen areas tended to be reduced above (23.57 mm(2)) and below (24.17 mm(2)) the stent. Lumen areas were consistent across stented regions at explantation (proximal stent 27.80 mm(2), mid stent 25.88 mm(2), and distal stent 25.81 mm(2)). The mean neointimal area in peritoneum-lined stents was 2.02±1.52 mm(2), with a neointima:media ratio of 1.03±0.50. These values were significantly lower in the above and below stent areas than in the stented regions, but there was no difference in either measure within the proximal, mid, or distal stent.

Conclusion: The custom-designed peritoneum-lined stent-graft is promising for clinical peripheral applications due to its ability to resist relevant long-term physiological stresses and outstanding short-term patency rates in canine implantations.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5621711PMC
http://dx.doi.org/10.1583/14-4762.1DOI Listing

Publication Analysis

Top Keywords

stent
11
fatigue vivo
8
axial/torsional fatigue
8
fatigue testing
8
peritoneum-lined stent-graft
8
proximal stent
8
mid stent
8
distal stent
8
lumen areas
8
stented regions
8

Similar Publications

Analysis of the effect of platelet function and different doses of ticagrelor after flow diverter treatment of intracranial aneurysms.

Neurosurg Rev

January 2025

Neurosurgery Center, Department of Cerebrovascular Surgery, Engineering Technology Research Center of Education Ministry of China on Diagnosis and Treatment of Cerebrovascular Disease, Zhujiang Hospital, Southern Medical University, Guangzhou, China.

Ticagrelor has become the standard drug for the treatment of intracranial aneurysms (IAs) with flow diverters (FDs), but the dosage has not been standardized. The effect of platelet function on clinical and imaging prognosis remains unclear. This study aimed to show the effects of different doses of ticagrelor and platelet aggregation function on the clinical and imaging prognosis after FDs treatment of aneurysms.

View Article and Find Full Text PDF

Background: Most patients undergoing pancreaticoduodenectomy (PD) for pancreatic ductal adenocarcinoma (PDAC) develop recurrence. No previous studies have investigated predictors of local-only recurrence following PD for PDAC. Our study aimed to determine timing, pattern and predictors of any-site and local-only recurrence following PD for PDAC.

View Article and Find Full Text PDF

Background: Early literature on the Woven EndoBridge (WEB) device reported 80-90% adequate aneurysm occlusion but low complete occlusion (40-55%). It is uncertain whether residual or recurrent aneurysms require re-treatment to prevent future rupture.

Objective: To systematically review the literature to meta-analyze occlusion and complication rates after re-treatment of these aneurysms.

View Article and Find Full Text PDF

Massive Dislodgement of an Infected Popliteal Stent Graft.

Eur J Vasc Endovasc Surg

January 2025

Vascular Surgery Unit, AOUP Policlinico "P. Giaccone", Palermo, Italy.

View Article and Find Full Text PDF

Rationale: The biodegradable polymer BioMatrix Alpha™ stent contains biolimus A9 drug which is sirolimus derivative increase in lipophicity. The biodegradable polymer sirolimus eluting Combo™ stent is a dual-therapy sirolimus-eluting and CD34+ antobody coated stent capturing endothelial progenitor cells (EPCs).

Hypothesis: The main hypothesis of the SORT OUT XI trial was that the biodegradable polymer biolimus A9 BioMatrix Alpha ™ stent is noninferior to the biodegradable polymer sirolimus eluting Combo™ stent in an all-comers population with coronary artery disease undergoing percutaneous coronary intervention (PCI).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!