Purpose: To investigate the use of a system using electromagnetic tracking (EMT), post-processing and an error-detection algorithm for detecting errors and resolving uncertainties in high-dose-rate brachytherapy catheter digitization for treatment planning.

Methods: EMT was used to localize 15 catheters inserted into a phantom using a stepwise acquisition technique. Five distinct acquisition experiments were performed. Noise associated with the acquisition was calculated. The dwell location configuration was extracted from the EMT data. A CT scan of the phantom was performed, and five distinct catheter digitization sessions were performed. No a priori registration of the CT scan coordinate system with the EMT coordinate system was performed. CT-based digitization was automatically extracted from the brachytherapy plan DICOM files (CT), and rigid registration was performed between EMT and CT dwell positions. EMT registration error was characterized in terms of the mean and maximum distance between corresponding EMT and CT dwell positions per catheter. An algorithm for error detection and identification was presented. Three types of errors were systematically simulated: swap of two catheter numbers, partial swap of catheter number identification for parts of the catheters (mix), and catheter-tip shift. Error-detection sensitivity (number of simulated scenarios correctly identified as containing an error/number of simulated scenarios containing an error) and specificity (number of scenarios correctly identified as not containing errors/number of correct scenarios) were calculated. Catheter identification sensitivity (number of catheters correctly identified as erroneous across all scenarios/number of erroneous catheters across all scenarios) and specificity (number of catheters correctly identified as correct across all scenarios/number of correct catheters across all scenarios) were calculated. The mean detected and identified shift was calculated.

Results: The maximum noise ±1 standard deviation associated with the EMT acquisitions was 1.0 ± 0.1 mm, and the mean noise was 0.6 ± 0.1 mm. Registration of all the EMT and CT dwell positions was associated with a mean catheter error of 0.6 ± 0.2 mm, a maximum catheter error of 0.9 ± 0.4 mm, a mean dwell error of 1.0 ± 0.3 mm, and a maximum dwell error of 1.3 ± 0.7 mm. Error detection and catheter identification sensitivity and specificity of 100% were observed for swap, mix and shift (≥2.6 mm for error detection; ≥2.7 mm for catheter identification) errors. A mean detected shift of 1.8 ± 0.4 mm and a mean identified shift of 1.9 ± 0.4 mm were observed.

Conclusions: Registration of the EMT dwell positions to the CT dwell positions was possible with a residual mean error per catheter of 0.6 ± 0.2 mm and a maximum error for any dwell of 1.3 ± 0.7 mm. These low residual registration errors show that quality assurance of the general characteristics of the catheters and of possible errors affecting one specific dwell position is possible. The sensitivity and specificity of the catheter digitization verification algorithm was 100% for swap and mix errors and for shifts ≥2.6 mm. On average, shifts ≥1.8 mm were detected, and shifts ≥1.9 mm were detected and identified.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7993398PMC
http://dx.doi.org/10.1118/1.4894710DOI Listing

Publication Analysis

Top Keywords

dwell positions
20
catheter digitization
16
emt dwell
16
correctly identified
16
catheter
13
error detection
12
catheter identification
12
error
11
emt
10
dwell
10

Similar Publications

Purpose: The present study evaluated the dosimetric impact and compared the dose variations between the advanced collapsed cone engine (Task Group 186) and Task Group 43 plans for cervical cancer using tandem and ovoid applicators.

Material And Methods: Thirty cervical cancer patients underwent iridium-192 (Ir) high-dose-rate (HDR) intra-cavitary brachytherapy using tandem and ovoid applicator. Original treatment plans for all patients were created using TG-43 dose calculation formalism.

View Article and Find Full Text PDF
Article Synopsis
  • - The study aimed to identify errors in transrectal ultrasound (TRUS)-based reconstructions for high-dose-rate brachytherapy in prostate cancer, using an electromagnetic tracking (EMT) system to enhance accuracy during treatment.
  • - Out of 265 evaluated needle reconstructions, 23% had minor errors or worse, with 9% classified as major or severe, primarily due to issues like incorrect needle placement and user errors.
  • - The findings indicate that one-quarter of the reconstructions had errors exceeding 2mm, highlighting the potential of EMT to help detect and prevent these mistakes, improving patient care without affecting the clinical process.
View Article and Find Full Text PDF

Contrast-enhanced CT is the standard imaging technique in oncological objectives. Rates of missed pathologies depend on work experience of the respective radiologists. Thus the aim of this study is to analyze the eye movements of professionals while reading CT images in order to evaluate whether the eye-fixation patterns and search strategies of experienced radiologists could explain higher detection rates of pathologies and whether such patterns can be learned.

View Article and Find Full Text PDF

Nucleosome flipping drives kinetic proofreading and processivity by SWR1.

Nature

December 2024

Single Molecule Biophysics Group, MRC Laboratory of Medical Sciences, London, UK.

The yeast SWR1 complex catalyses the exchange of histone H2A-H2B dimers in nucleosomes, with Htz1-H2B dimers. Here we used single-molecule analysis to demonstrate two-step double exchange of the two H2A-H2B dimers in a canonical yeast nucleosome with Htz1-H2B dimers, and showed that double exchange can be processive without release of the nucleosome from the SWR1 complex. Further analysis showed that bound nucleosomes flip between two states, with each presenting a different face, and hence histone dimer, to SWR1.

View Article and Find Full Text PDF

Background: Comprehensive quality assurance (QA) for a seamless workflow of high-dose-rate brachytherapy, from imaging to planning and irradiation, is uncommon, and QA of the source dwell position is performed in one- or two-dimensions. Gel dosimetry using magnetic resonance imaging (MRI) is effective in verifying the three-dimensional distribution of doses for image-guided brachytherapy (IGBT). However, MRI scanners are not readily accessible, and MRI scanning is time-consuming.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!