Validation of a model-based measurement of the minimum insert thickness of knee prostheses: a retrieval study.

Bone Joint Res

Leiden University Medical Center, Department of Orthopaedics, Biomechanics and Imaging Group, PO 9600, 2300 RC, Leiden, The Netherlands.

Published: October 2014

Introduction: Wear of polyethylene inserts plays an important role in failure of total knee replacement and can be monitored in vivo by measuring the minimum joint space width in anteroposterior radiographs. The objective of this retrospective cross-sectional study was to compare the accuracy and precision of a new model-based method with the conventional method by analysing the difference between the minimum joint space width measurements and the actual thickness of retrieved polyethylene tibial inserts.

Method: Before revision, the minimum joint space width values and their locations on the insert were measured in 15 fully weight-bearing radiographs. These measurements were compared with the actual minimum thickness values and locations of the retrieved tibial inserts after revision.

Results: The mean error in the model-based minimum joint space width measurement was significantly smaller than the conventional method for medial condyles (0.50 vs 0.94 mm, p < 0.01) and for lateral condyles (0.06 vs 0.34 mm, p = 0.02). The precision (standard deviation of the error) of the methods was similar (0.84 vs 0.79 mm medially and both 0.46 mm laterally). The distance between the true minimum joint space width locations and the locations from the model-based measurements was less than 10 mm in the medial direction in 12 cases and less in the lateral direction in 13 cases.

Conclusion: The model-based minimum joint space width measurement method is more accurate than the conventional measurement with the same precision. Cite this article: Bone Joint Res 2014;3:289-96.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4220171PMC
http://dx.doi.org/10.1302/2046-3758.310.2000304DOI Listing

Publication Analysis

Top Keywords

minimum joint
24
joint space
24
space width
24
minimum
8
conventional method
8
values locations
8
model-based minimum
8
width measurement
8
joint
7
space
6

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!