Silicon photodiodes used for soft x-ray detection typically have a thin metal electrode partially covering the active area of the photodiode, which subtly alters the spectral sensitivity of the photodiode. As a specific example, AXUV4BST photodiodes from International Radiation Detectors have a 1.0 μm thick aluminum frame covering 19% of the active area of the photodiode, which attenuates the measured x-ray signal below ~6 keV. This effect has a small systematic impact on the electron temperature calculated from measurements of soft x-ray bremsstrahlung emission from a high-temperature plasma. Although the systematic error introduced by the aluminum frame is only a few percent in typical experimental conditions on the Madison Symmetric Torus, it may be more significant for other instruments that use similar detectors.

Download full-text PDF

Source
http://dx.doi.org/10.1063/1.4894656DOI Listing

Publication Analysis

Top Keywords

soft x-ray
12
spectral sensitivity
8
active area
8
area photodiode
8
aluminum frame
8
note photodiode
4
photodiode aluminum
4
aluminum cathode
4
cathode frame
4
frame spectral
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!