AI Article Synopsis

  • The study aimed to compare intraocular pressure (IOP) measurements from the Icare Pro and Tono-Pen XL tonometers while subjects were in different body positions, such as seated, supine, and lateral decubitus.
  • A total of 100 eyes from 50 participants were evaluated, showing that while there was good correlation between the two devices in measuring IOP, the Icare Pro consistently reported higher readings than the Tono-Pen across all positions.
  • The differences in IOP readings were found to be weakly correlated with central corneal thickness, particularly in the dependent eye when lying on one side, highlighting the importance of considering tonometer choice and body position in clinical assessments of IOP.

Article Abstract

Purpose: To compare intraocular pressure (IOP) measurements obtained using the Icare Pro rebound tonometer and Tono-Pen XL tonometer in supine and lateral decubitus body positions.

Methods: One-hundred eyes of 50 subjects (normal volunteers or glaucoma suspects) were enrolled in this prospective observational study. IOP was measured in both eyes using the Icare Pro and Tono-Pen XL in the sitting position and the recumbent positions including supine, right lateral decubitus and left lateral decubitus. IOP was measured five minutes after assuming each of the recumbent postures in a randomized sequence. The eye on the lower side in the lateral decubitus position was termed as the dependent eye. Agreement of IOP readings between the Icare Pro and Tono-Pen was assessed in all recumbent positions. Differences of IOP readings (ΔIOP) between the two tonometers and their correlations with ocular parameters were also assessed in all positions.

Results: The IOP readings obtained using Icare Pro and Tono-Pen showed good correlations in supine and lateral decubitus positions (all r > 0.7, p < 0.005), although Icare Pro readings were higher than Tono-Pen readings (all p < 0.001) in all positions. The ΔIOP showed a weakly positive correlation with central corneal thickness in both eyes, whereas such a positive correlation was found only in the dependent eye in the lateral decubitus positions (r = 0.307-0.531, all p < 0.005). Both the spherical equivalents and axial lengths were not correlated with ΔIOP in all positions.

Conclusion: IOP readings obtained with Tono-Pen and Icare Pro tonometers showed good agreement in supine as well as in lateral decubitus positions, although Icare readings were higher than Tono-Pen readings in all positions. Such differences in IOP readings between the different tonometers need to be considered when measuring IOPs in various body positions.

Download full-text PDF

Source
http://dx.doi.org/10.3109/02713683.2014.964416DOI Listing

Publication Analysis

Top Keywords

lateral decubitus
24
icare pro
20
supine lateral
16
pro tono-pen
12
iop readings
12
intraocular pressure
8
measurements icare
8
pro rebound
8
rebound tonometer
8
tonometer tono-pen
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!