Introduction: The combination of the LUCAS 2 (L-CPR) automated CPR device and an impedance threshold device (ITD) has been widely implemented in the clinical field. This animal study tested the hypothesis that the addition of an ITD on L-CPR would enhance cerebral and coronary perfusion pressures.

Methods: Ten female pigs (39.0 ± 2.0 kg) were sedated, intubated, anesthetized with isofluorane, and paralyzed with succinylcholine (93.3 μg/kg/min) to inhibit the potential confounding effect of gasping. After 4 min of untreated ventricular fibrillation, 4 min of L-CPR+an active ITD or L-CPR+a sham ITD was initiated and followed by another 4 min of the alternative method of CPR. Systolic blood pressure (SBP), diastolic blood pressure (DBP), diastolic right atrial pressure (RAP), intracranial pressure (ICP), airway pressure, and end tidal CO2 (ETCO2) were recorded continuously. Data expressed as mean mmHg ± SD.

Results: Decompression phase airway pressure was significantly lower with L-CPR+active ITD versus L-CPR+sham ITD (-5.3 ± 2.2 vs. -0.5 ± 0.6; p<0.001). L-CPR+active ITD treatment resulted in significantly improved hemodynamics versus L-CPR+sham ITD: ETCO2, 35 ± 6 vs. 29 ± 7 (p=0.015); SBP, 99 ± 9 vs. 93 ± 15 (p=0.050); DBP, 24 ± 12 vs. 19 ± 15 (p=0.006); coronary perfusion pressure, 29 ± 8 vs. 26 ± 7 (p=0.004) and cerebral perfusion pressure, 24 ± 13 vs. 21 ± 12 (p=0.028).

Conclusions: In pigs undergoing L-CPR the addition of the active ITD significantly reduced intrathoracic pressure and increased vital organ perfusion pressures.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.resuscitation.2014.09.013DOI Listing

Publication Analysis

Top Keywords

impedance threshold
8
threshold device
8
blood pressure
8
airway pressure
8
pressure
6
hemodynamic improvement
4
improvement lucas
4
lucas automated
4
device
4
automated device
4

Similar Publications

A long-term clinical comparative study of left bundle branch pacing versus biventricular pacing in patients with heart failure and complete left bundle branch block.

Heart Vessels

January 2025

Department of Cardiology, Fujian Medical University Union Hospital, Fujian Institute of Coronary Heart Disease, Fujian Heart Medical Center, Fuzhou, 350001, Fujian, China.

Left bundle branch pacing (LBBP) is an emerging physiological pacing technique characterized by stable pacing parameters and a narrower QRS duration. This study aims to compare the long-term efficacy and safety of biventricular pacing (BIVP) and LBBP in patients with heart failure with reduced ejection fraction (HFrEF) and complete left bundle branch block (CLBBB). A retrospective analysis was conducted on 35 patients with chronic HFrEF accompanied by CLBBB treated at our center from April 2018 to October 2022.

View Article and Find Full Text PDF

Background: According to the Lyon Consensus 2.0, mean nocturnal baseline impedance (MNBI) greater than 2500 Ω is considered evidence against gastroesophageal reflux disease (GERD). However, we have observed that a subset of GERD patients presents with MNBI exceeding 2500 Ω.

View Article and Find Full Text PDF

Background: Noninferiority of omitting intraoperative defibrillation threshold (DFT) testing has been documented for transvenous implantable cardioverter defibrillators (ICD) whereas data for the subcutaneous-ICD (S-ICD) regarding the need for DFT testing, especially during S-ICD generator replacement, is not available.

Methods: A total of 112 consecutive patients who underwent S-ICD generator replacement and routine testing were included in this retrospective single-center study and analyzed regarding the outcome of intraoperative DFT.

Results: The majority of patients (87.

View Article and Find Full Text PDF

Biological relevance and methodological implications of unexpected hearing thresholds in a diving bird.

Sci Rep

December 2024

Deutsches Meeresmuseum, Katharinenberg 14 - 20, 18439, Stralsund, Germany.

Many animals alternate between different media, such as air and water, thanks to specific adaptations. Among birds, penguins (Sphenisciformes) have the most extreme morphological, physiological, and behavioural adaptations to their amphibious lifestyle. Their auditory perception of sound, potentially matching different impedances in air and under water, is largely unknown particularly in terms of whether their underwater adaptations may have affected their in-air hearing capacity.

View Article and Find Full Text PDF

A wearable triboelectric impedance tomography system for noninvasive and dynamic imaging of biological tissues.

Sci Adv

December 2024

Beijing Key Laboratory of Micro-Nano Energy and Sensor, Center for High-Entropy Energy and Systems, Beijing Institute of Nanoenergy and Nanosystems, Chinese Academy of Sciences, Beijing 101400, P. R. China.

Tissue imaging is usually captured by hospital-based nuclear magnetic resonance. Here, we present a wearable triboelectric impedance tomography (TIT) system for noninvasive imaging of various biological tissues. The imaging mechanism relies on the obtained impedance information from the different soft human tissues.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!