Background: We performed a multicenter evaluation of the Elecsys® progastrin-releasing peptide (ProGRP) immunoassay in Europe and China.

Methods: The assay was evaluated at three European and two Chinese sites by imprecision, stability, method comparison and differentiation potential in lung cancer.

Results: Intermediate imprecision across five analyte concentrations ranged from 2.2% to 6.0% coefficient of variation. Good stability for plasma and serum samples was shown for various storage conditions. There was excellent correlation between the Elecsys® and ARCHITECT assays in plasma (slope 1.02, intercept -2.72pg/mL). The Elecsys® assay also showed good correlation between serum and plasma samples (slope 0.93, intercept 2.35pg/mL; correlation coefficient 0.97). ProGRP differentiated small-cell and non-small-cell lung cancer (NSCLC; area under the curve 0.90, 95% CI 0.87-0.93; 78.3% sensitivity, 95% specificity; at 84pg/mL), with no relevant effects of ethnicity, age, gender or smoking. Median ProGRP concentrations were low in benign diseases (38pg/mL), other malignancies (40pg/mL) or NSCLC (39pg/mL), except chronic kidney disease above stage 3 (>100pg/mL).

Conclusions: Increased stability of the Elecsys® ProGRP assay in serum and plasma offers clear benefits over existing assays. This first evaluation of a ProGRP assay in China demonstrated comparable differentiation potential among different ethnicities.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.cca.2014.09.015DOI Listing

Publication Analysis

Top Keywords

multicenter evaluation
8
progastrin-releasing peptide
8
peptide progrp
8
progrp immunoassay
8
immunoassay europe
8
differentiation potential
8
serum plasma
8
progrp assay
8
progrp
6
evaluation progastrin-releasing
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!