The development of computer-based 3D models of the aortic root is one of the most important problems in constructing the prostheses for transcatheter aortic valve implantation. In the current study, we analyzed data from 117 patients with and without aortic valve disease and computed tomography data from 20 patients without aortic valvular diseases in order to estimate the average values of the diameter of the aortic annulus and other aortic root parameters. Based on these data, we developed a 3D model of human aortic root with unique geometry. Furthermore, in this study we show that by applying different material properties to the aortic annulus zone in our model, we can significantly improve the quality of the results of finite element analysis. To summarize, here we present four 3D models of human aortic root with unique geometry based on computational analysis of ECHO and CT data. We suggest that our models can be utilized for the development of better prostheses for transcatheter aortic valve implantation.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.compbiomed.2014.08.023DOI Listing

Publication Analysis

Top Keywords

aortic root
20
human aortic
12
aortic valve
12
aortic
11
prostheses transcatheter
8
transcatheter aortic
8
valve implantation
8
patients aortic
8
aortic annulus
8
root unique
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?