Background: Transcatheter aortic valve implantation involves percutaneously implanting a biomechanical aortic valve to treat severe aortic stenosis. In order to select a proper device, precise sizing of the aortic valve annulus must be completed.

Methods: In this paper, we describe a fully automatic segmentation method to measure the aortic annulus diameter in patients with aortic calcification, operating on 3-dimensional transesophageal echocardiographic images. The method is based on state estimation of a subdivision surface representation of the left ventricular outflow tract and aortic root. The state estimation is solved by an extended Kalman filter driven by edge detections normal to the subdivision surface.

Results: The method was validated on echocardiographic recordings of 16 patients. Comparison against two manual measurements showed agreements (mean ±SD) of -0.3 ± 1.6 and -0.2 ± 2.3 mm for perimeter-derived diameters, compared to an interobserver agreement of -0.1 ± 2.1 mm.

Conclusions: With this study, we demonstrated the feasibility of an efficient and fully automatic measurement of the aortic annulus in patients with aortic disease. The algorithm robustly measured the aortic annulus diameter, providing measurements indistinguishable from those done by cardiologists.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4274687PMC
http://dx.doi.org/10.1186/1471-2342-14-31DOI Listing

Publication Analysis

Top Keywords

aortic annulus
16
annulus diameter
12
aortic valve
12
aortic
11
automatic measurement
8
measurement aortic
8
fully automatic
8
patients aortic
8
state estimation
8
annulus
5

Similar Publications

Severe mitral regurgitation (MR) following surgical repair of the mitral valve poses a significant clinical challenge. Patients who have undergone surgery are typically at high risk for a second operation. This report details the case of a 54-year-old male who underwent aortic valve replacement and mitral valve repair using a 34-ring, 14 years prior.

View Article and Find Full Text PDF

The Y-incision aortic annular enlargement (AAE) has been established as a safe and effective technique for upsizing the aortic annulus by 3 to 4 valve sizes. However, concerns have been raised regarding its technical complexity during reoperations, particularly given the extensive enlargement of the aortic annulus and root. We present a case of reoperative aortic valve replacement after previous Y-incision AAE for prosthetic valve endocarditis and aortic root abscess.

View Article and Find Full Text PDF

Among repairs for ventriculoarterial discordance, ventricular septal defect, and pulmonary stenosis, aortic root translocation (Nikaidoh operation) offers the most anatomic result. With a diminutive pulmonary annulus or hypoplastic left ventricular outflow tract, the distance gained posteriorly with aortic translocation is negligible. We developed the "hemi-Nikaidoh" procedure as an alternative.

View Article and Find Full Text PDF

Background: Partial heart transplantation (PHT) is a novel procedure for children in need of a growing valve replacement option. One challenge is identifying suitable donor valves. Semilunar heart valves from patients receiving a retransplant may be a source, however their functionality and growth potential especially at the time of retransplant are unknown.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!