Very low-dose (0.15 mGy) chest CT protocols using the COPDGene 2 test object and a third-generation dual-source CT scanner with corresponding third-generation iterative reconstruction software.

Invest Radiol

From the Departments of *Radiology, and †Biomedical Engineering, University of Iowa, Iowa City, IA; ‡Siemens Healthcare, Erlangen; §Siemens Healthcare AG, Forchheim, Germany; Departments of ∥Statistics and Actuarial Science, and ¶Medicine, University of Iowa, Iowa City, IA.

Published: January 2015

Objectives: The purpose of this study was to evaluate the impact of ultralow radiation dose single-energy computed tomographic (CT) acquisitions with Sn prefiltration and third-generation iterative reconstruction on density-based quantitative measures of growing interest in phenotyping pulmonary disease.

Materials And Methods: The effects of both decreasing dose and different body habitus on the accuracy of the mean CT attenuation measurements and the level of image noise (SD) were evaluated using the COPDGene 2 test object, containing 8 different materials of interest ranging from air to acrylic and including various density foams. A third-generation dual-source multidetector CT scanner (Siemens SOMATOM FORCE; Siemens Healthcare AG, Erlangen, Germany) running advanced modeled iterative reconstruction (ADMIRE) software (Siemens Healthcare AG) was used.We used normal and very large body habitus rings at dose levels varying from 1.5 to 0.15 mGy using a spectral-shaped (0.6-mm Sn) tube output of 100 kV(p). Three CT scans were obtained at each dose level using both rings. Regions of interest for each material in the test object scans were automatically extracted. The Hounsfield unit values of each material using weighted filtered back projection (WFBP) at 1.5 mGy was used as the reference value to evaluate shifts in CT attenuation at lower dose levels using either WFBP or ADMIRE. Statistical analysis included basic statistics, Welch t tests, multivariable covariant model using the F test to assess the significance of the explanatory (independent) variables on the response (dependent) variable, and CT mean attenuation, in the multivariable covariant model including reconstruction method.

Results: Multivariable regression analysis of the mean CT attenuation values showed a significant difference with decreasing dose between ADMIRE and WFBP. The ADMIRE has reduced noise and more stable CT attenuation compared with WFBP. There was a strong effect on the mean CT attenuation values of the scanned materials for ring size (P < 0.0001) and dose level (P < 0.0001). The number of voxels in the region of interest for the particular material studied did not demonstrate a significant effect (P > 0.05). The SD was lower with ADMIRE compared with WFBP at all dose levels and ring sizes (P < 0.05).

Conclusions: The third-generation dual-source CT scanners using third-generation iterative reconstruction methods can acquire accurate quantitative CT images with acceptable image noise at very low-dose levels (0.15 mGy). This opens up new diagnostic and research opportunities in CT phenotyping of the lung for developing new treatments and increased understanding of pulmonary disease.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4294320PMC
http://dx.doi.org/10.1097/RLI.0000000000000093DOI Listing

Publication Analysis

Top Keywords

iterative reconstruction
16
015 mgy
12
test object
12
third-generation dual-source
12
third-generation iterative
12
dose levels
12
copdgene test
8
dose
8
decreasing dose
8
body habitus
8

Similar Publications

Background: Deep learning image reconstruction (DLIR) algorithms allow strong noise reduction while preserving noise texture, which may potentially improve hypervascular focal liver lesions.

Purpose: To assess the impact of DLIR on image quality (IQ) and detectability of simulated hypervascular hepatocellular carcinoma (HCC) in fast kV-switching dual-energy CT (DECT).

Methods: An anthropomorphic phantom of a standard patient morphology (body mass index of 23 kg m) with customized liver, including mimickers of hypervascular lesions in both late arterial phase (AP) and portal venous phase (PVP) enhancement, was scanned on a DECT.

View Article and Find Full Text PDF

Background And Purpose:  The amount of information publicly available from arthroplasty registries is large but could be used more effectively. This project aims to improve the knowledge concerning existing registries to facilitate access, transparency, harmonization, and reporting.

Methods: Within the International Society of Arthroplasty Registries (ISAR) we aimed at developing, testing, adopting, and making publicly available a short, standardized registry description with items considered relevant for stakeholders using a cross-sectional study survey.

View Article and Find Full Text PDF

We describe a novel application of photon-counting detector CT (PCD-CT) in neurovascular imaging by harnessing the improved spatial resolution, attenuation of electronic noise, and reduction of metal artifacts. The presented case offers the unique challenge of high-quality imaging for the assessment of treated and untreated intracranial saccular aneurysms, in the setting of metal artifacts from embolization coils. Our goal was to explore optimized reconstruction parameters for ultra-high-resolution imaging (UHR) using a dedicated, sharp neurovascular kernel (Hv72) and the highest strength of quantum iterative reconstruction (QIR-4) for detailed characterization of the vasculature.

View Article and Find Full Text PDF

Background: With the widespread use of lumbar pedicle screws for internal fixation, the morphology of the screws and the surrounding tissues should be evaluated. The metal artifact reduction (MAR) technique can reduce the artifacts caused by pedicle screws, improve the quality of computed tomography (CT) images after pedicle fixation, and provide more imaging information to the clinic.

Purpose: To explore whether the MAR+ method, a projection-based algorithm for correcting metal artifacts through multiple iterative operations, can reduce metal artifacts and have an impact on the structure of the surrounding metal.

View Article and Find Full Text PDF

Synchrotron X-ray microtomography (S-µCT) is a highly valuable technique for investigating organ function and pathologies. However, its application is often limited by high radiation doses and the occurrence of ring artifacts. While S-µCT utilizing sparse-view projections can effectively decrease radiation doses, the reconstructed images frequently exhibit severe streaking artifacts, which are exacerbated by ring artifacts, ultimately compromising reconstruction accuracy, image quality, and resolution.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!