Surface plasmon field-enhanced fluorescence spectroscopy (SPFS) is a promising methodology for point-of-care (POC) testing. The SPFS devices that have been reported are equipped with an angle rotating stage to adjust the surface plasmon resonance (SPR) angle. In a clinical setting, however, the SPR angle determination is a tedious and time-consuming process. In this study, we employed an SPFS instrument with a convergent optical system that allows the omission of this procedure. We demonstrated that this instrumentation allowed the sensitive determination of low concentrations of α-fetoprotein in serum and reduced the variation effect caused by the protein concentrations in samples. The SPFS with a convergent optical system is suitable for POC testing.