Effect of fluid contamination on reverse torque values in bone-level implants.

Implant Dent

*Assistant Professor, Department of Prosthodontics, Faculty of Dentistry, Erciyes University, Kayseri, Turkey. †Associate Professor, Department of Prosthodontics, Faculty of Dentistry, Erciyes University, Kayseri, Turkey. ‡Research Assistant, Department of Prosthodontics, Faculty of Dentistry, Erciyes University, Kayseri, Turkey. §Assistant Professor, Department of Prosthodontics, Faculty of Dentistry, Pamukkale University, Denizli, Turkey.

Published: October 2014

Introduction: Screw loosening is one of the most common mechanical complications within the prosthetic components of dental implants. Clinicians must know if reverse torque values (RTVs) are affected by contamination of the implant abutment screw hole with either blood, saliva or chlorhexidine.

Methods: Forty implants were divided into 4 groups (n = 10). Implants were embedded in epoxy resin and placed in a digital torque meter. Screw holes of 30 implants were contaminated with either chlorhexidine (CG), fresh human saliva (SG), or blood (BG), and the remaining noncontaminated screws served as a control group (NC). Abutment screws were tightened using a hand-held torque meter (25 N·cm). Thermocycling (1500 cycles) was performed, and abutment detorque values were calculated using the digital torque meter.

Results: RTVs were lower than insertion torque values for all groups. According to Tukey HSD Test, blood contamination of abutment screws resulted in significantly lower RTVs (18.950 ± 2.641 N·cm) when compared with controls.

Conclusions: Blood contamination of abutment screw holes was shown to significantly decrease RTVs. Blood contamination may result in greater loosening of implant screws in clinical practice.

Download full-text PDF

Source
http://dx.doi.org/10.1097/ID.0000000000000139DOI Listing

Publication Analysis

Top Keywords

torque values
12
blood contamination
12
reverse torque
8
abutment screw
8
digital torque
8
torque meter
8
screw holes
8
abutment screws
8
contamination abutment
8
torque
6

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!