Purpose: 90Y-microspheres are widely used for the radioembolization of metastatic liver cancer or hepatocellular carcinoma and there is a growing interest for imaging 90Y-microspheres with PET. The aim of this study is to evaluate the performance of a current generation PET/CT scanner for 90Y imaging and to optimize the PET protocol to improve the assessment and the quantification of 90Y-microsphere biodistribution after radioembolization.

Methods: Data were acquired on a Biograph mCT-TrueV scanner with time of flight (TOF) and point spread function (PSF) modeling. Spatial resolution was measured with a 90Y point source. Sensitivity was evaluated using the NEMA 70 cm line source filled with 90Y. To evaluate the count rate performance, 90Y vials with activity ranging from 3.64 to 0.035 GBq were measured in the center of the field of view (CFOV). The energy spectrum was evaluated. Image quality with different reconstructions was studied using the Jaszczak phantom containing six hollow spheres (diameters: 31.3, 28.1, 21.8, 16.1, 13.3, and 10.5 mm), filled with a 207 kBq/ml 90Y concentration and a 5:1 sphere-to-background ratio. Acquisition time was adjusted to simulate the quality of a realistic clinical PET acquisition of a patient treated with SIR-Spheres®. The developed methodology was applied to ten patients after SIR-Spheres® treatment acquiring a 10 min per bed PET.

Results: The energy spectrum showed the 90Y bremsstrahlung radiation. The 90Y transverse resolution, with filtered backprojection reconstruction, was 4.5 mm in the CFOV and degraded to 5.0 mm at 10 cm off-axis. 90Y absolute sensitivity was 0.40 kcps/MBq in the center of the field of view. Tendency of true and random rates as a function of the 90Y activity could be accurately described using linear and quadratic models, respectively. Phantom studies demonstrated that, due to low count statistics in 90Y PET acquisition, the optimal parameters for the standard OSEM+PSF reconstruction were only one iteration and a postreconstruction filter of 6 mm FWHM, for both TOF and non-TOF reconstructions. Moreover, when TOF is included, the signal to noise ratio increased and visibility achieved 100% by the experienced observers and 93.3% according to the Rose model of statistical detection. In 50% of patients, TOF allowed the visualization of 90Y radioembolized lesions not seen without TOF, confirming phantom results. Liver activity was accurately quantified, with no significant differences between reconstructed and actual delivered activity to the whole-liver [mean relative difference (10.2±14.7)%].

Conclusions: Qualitative and quantitative 90Y PET imaging improved with the introduction of TOF in a PET/CT scanner, thereby allowing the visualization of microsphere deposition in lesions not visible in non-TOF images. This technique accurately quantifies the total activity delivered to the liver during radioembolization with (90)Y-microspheres and allows dose estimation.

Download full-text PDF

Source
http://dx.doi.org/10.1118/1.4892383DOI Listing

Publication Analysis

Top Keywords

90y
12
quantification 90y-microsphere
8
90y-microsphere biodistribution
8
pet/ct scanner
8
center field
8
field view
8
energy spectrum
8
pet acquisition
8
activity accurately
8
90y pet
8

Similar Publications

In this study, Bayesian linear regression was employed to simultaneously determine Cherenkov counting efficiencies for ⁸⁹Sr, ⁹⁰Sr, and ⁹⁰Y from a mixed standard. Among the two codes utilized, PyMC3 and Stan, the latter produced more reliable posterior distributions for the three efficiencies, particularly for ⁹⁰Sr which is challenging to estimate due to its minimal activity changes. Validation showed excellent agreement between the activities determined with Stan's estimates and the certified activities, implying the applicability of the Bayesian approach.

View Article and Find Full Text PDF

Background: Yttrium-90 FF-21101 (Y-FF-21101) is a radiopharmaceutical that targets P-cadherin as a therapy against solid tumors. A previously reported, first-in-human study determined that a dose of 25 mCi/m was safe, and a patient with clear cell carcinoma of the ovary achieved a complete response. In this article, the authors report the results of Y-FF-21101 treatment in an ovarian carcinoma expansion cohort and in patients with selected solid tumors who had known high P-cadherin expression.

View Article and Find Full Text PDF

Purpose: Long axial field-of-view (LAFOV) positron emission tomography/computed tomography (PET/CT) scanners enable high sensitivity and wide anatomical coverage. Therefore, they seem ideal to perform post-selective internal radiation therapy (SIRT) Y scans, which are needed, to confirm that the dose is delivered to the tumors and that healthy organs are spared. However, it is unclear to what extent the use of LAFOV PET is feasible and which dosimetry approaches results in accurate measurements.

View Article and Find Full Text PDF

In same-day radioembolization, 99mTc-MAA SPECT/CT, 90Y radioembolization, and post-treatment 90Y SPECT/CT procedures are conducted on the same-day, resulting in a dual-isotope environment of 90Y and 99mTc during post-treatment imaging. This study aimed to quantify the impact of 99mTc on 90Y bremsstrahlung-SPECT/CT image quality and to establish an optimised imaging protocol for both clinical practice, and with advanced reconstruction techniques. Utilising a NEMA IQ phantom, contrast recovery coefficients (CRCs) were measured to evaluate the 90Y image quality degradation caused by 99mTc.

View Article and Find Full Text PDF

Background: In radioembolization therapy for hepatic malignancies, the accurate estimation of lung shunt fraction (LSF) is crucial to minimize the risk of radiation-induced pneumonitis and fibrosis due to hepatopulmonary shunting of yttrium-90 (90Y)-microspheres. This study aimed to compare the accuracy and precision of LSF estimation using technetium-99m macroaggregated albumin single photon emission computed tomography ([99mTc]Tc-MAA SPECT) LSF, [99mTc]Tc-MAA planar LSF, and 90Y PET LSF in patients undergoing 90Y-radioembolization.

Material And Methods: A retrospective study was conducted involving 15 patients diagnosed with hepatocellular carcinoma (HCC) or liver metastases and planned to undergo transarterial radioembolization with 90Y SirSpheres after multidisplinary team discussion.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!