Evaluation of a transcutaneous blood gas monitoring system in critically ill dogs.

J Vet Emerg Crit Care (San Antonio)

Department of Clinical Studies, Ontario Veterinary College, University of Guelph, 50 Stone Road East, Guelph, ON, N1G 2W1, Canada.

Published: June 2015

Objectives: To describe the use of a transcutaneous blood gas monitoring system in critically ill dogs, determine if transcutaneous and arterial blood gas values have good agreement, and verify if clinical or laboratory variables are correlated with differences between transcutaneous and arterial blood gas measurements.

Design: Prospective observational study.

Setting: University teaching hospital ICU.

Animals: Twenty-three client-owned dogs.

Interventions: In critically ill dogs undergoing arterial blood gas monitoring, a transcutaneous blood gas monitor was used to measure transcutaneous partial pressure of carbon dioxide (PtcCO2 ) and transcutaneous partial pressure of oxygen (PtcO2 ) values 30 minutes after sensor placement, which were compared to PaCO2 and PaO2 values measured simultaneously. Clinical and laboratory variables were concurrently recorded to determine if they were correlated with the difference between transcutaneous and arterial blood gas measurements.

Measurements And Main Results: Bland-Altman analysis revealed a mean bias of 4.6 ± 26.3 mm Hg (limits of agreement [LOA]: -46.9/+56.1 mm Hg) between PtcO2 and PaO2 and a mean bias of 9.3 ± 8.5 mm Hg (LOA: -7.5/+26.0 mm Hg) between PtcCO2 and PaCO2 . The difference between PtcCO2 -PaCO2 was strongly negatively correlated with HCO3 (-) (r(2) = 0.52, P < 0.001) and PaCO2 (r(2) = 0.58, P < 0.001) and weakly positively correlated with diastolic blood pressure (r(2) = 0.21, P = 0.044), whereas the difference between PtcCO2 -PaCO2 was moderately negatively correlated with diastolic blood pressure (r(2) = 0.33, P = 0.008).

Conclusions: Agreement between transcutaneous and arterial PO2 and PCO2 measurements in these critically ill dogs was inferior to that reported in similar adult and pediatric human studies. The transcutaneous monitor consistently over-estimated PaO2 and PaCO2 and should not be used to replace arterial blood gas measurements in critically ill dogs requiring blood gas interpretation.

Download full-text PDF

Source
http://dx.doi.org/10.1111/vec.12216DOI Listing

Publication Analysis

Top Keywords

blood gas
36
critically ill
20
ill dogs
20
arterial blood
20
transcutaneous arterial
16
transcutaneous blood
12
gas monitoring
12
blood
11
gas
9
transcutaneous
9

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!