Purpose: To evaluate, first, the feasibility of using Pictor, a new portable, noncontact digital fundus camera, to obtain high-quality retinal images of prematurely born infants and, second, the accuracy of grading these images for clinically significant posterior pole vascular changes, that is, pre-plus or plus disease, compared to indirect ophthalmoscopy.

Methods: Pictor retinal images were obtained on a convenience sample of prematurely born infants during routine examinations for retinopathy of prematurity (ROP). Images were reviewed and graded by two ROP experts, who were masked to demographic and clinical examination findings.

Results: A total of 96 eyes of 48 infants were included. The mean field of view was 5.0 × 6.1 disk diameters (DD). Grader 1 found quality to be fair or good in 96% of images; grader 2 in 97% of images. Grader 1 judged images as having at least 1 DD length of a major vessel in 3 or 4 quadrants in 80% of images; grader 2 in 86% of images. The sensitivity and specificity of grading pre-plus or plus disease on Pictor images was 100% and 79%, respectively, for grader 1, and 83% and 85%, respectively, for grader 2, compared to the reference standard of indirect ophthalmoscopy.

Conclusions: The Pictor fundus camera can capture digital retinal images of prematurely born infants that have quality sufficient for accurate expert grading in comparison to clinical examination. Pictor shows promise as an ROP screening tool.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4150087PMC
http://dx.doi.org/10.1016/j.jaapos.2014.02.013DOI Listing

Publication Analysis

Top Keywords

retinal images
12
prematurely born
12
born infants
12
images grader
12
images
11
noncontact digital
8
fundus camera
8
images prematurely
8
pre-plus disease
8
clinical examination
8

Similar Publications

Comparison of Ultrasound Characteristics of Peripapillary Hyperreflective Ovoid Mass-Like Structures (PHOMS) and Optic Disc Drusen in Children.

J Neuroophthalmol

January 2025

Departments of Ophthalmology (DB, G-SY, GTL, RAA) and Neurology (DB, GTL, RAA), Perelman School of Medicine at the University of Pennsylvania, Philadelphia, Pennsylvania; and Division of Ophthalmology (AG, GTL, RAA), Children's Hospital of Philadelphia, Philadelphia, Pennsylvania.

Background: In children, pseudopapilledema is frequently caused by peripapillary hyperreflective ovoid mass-like structures (PHOMS) or optic disc drusen (ODD). While enhanced depth imaging (EDI) OCT can identify both, lack of cooperation, especially from younger children due to the duration of testing, often necessitates the use of B-scan ultrasound (BSUS). This study investigated whether PHOMS are hyperreflective on BSUS and if BSUS can differentiate PHOMS from ODD.

View Article and Find Full Text PDF

Purpose: To report a case of biopsy-proven sarcoidosis in a patient with panuveitis and a positive interferon-gamma release assay (IGRA) from a non-endemic tuberculosis (TB) country.

Methods: Case report.

Results: A 26-year-old male from the United Arab Emirates (UAE) presented with granulomatous panuveitis characterized by mutton-fat keratic precipitates, anterior chamber and vitreous cells, and retinal vasculitis.

View Article and Find Full Text PDF

To assess the repeatability of a microperimetry methodology for quantifying visual function changes in the junctional zone of eyes with geographic atrophy (GA) in the clinical trial context. A post hoc analysis of the OAKS phase III trial was conducted, which enrolled patients with GA secondary to age-related macular degeneration. Microperimetry using a standard 10-2 fovea centered grid was performed at baseline and follow-up visits.

View Article and Find Full Text PDF

Adaptive optics scanning light ophthalmoscopy (AOSLO) enables high-resolution retinal imaging, eye tracking, and stimulus delivery in the living eye. AOSLO-mediated visual stimuli are created by temporally modulating the excitation light as it scans across the retina. As a result, each location within the field of view receives a brief flash of light during each scanner cycle (every 33-40 ms).

View Article and Find Full Text PDF

Diverse retinal ganglion cells (RGCs) transmit distinct visual features from the eye to the brain. Recent studies have categorized RGCs into 45 types in mice based on transcriptomic profiles, showing strong alignment with morphological and electrophysiological properties. However, little is known about how these types are spatially arranged on the two-dimensional retinal surface-an organization that influences visual encoding-and how their local microenvironments impact development and neurodegenerative responses.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!