The identification and quantification of a high molecular weight light stabilizer (Uvinul 3030™) in an unknown polycarbonate sample was achieved through the application of SEC-Py-TOF-GCMS. A size exclusion column optimized to achieve resolution in the lower mass range was applied to allow the fractionation of an individual additive peak. A commercially available sampling interface was operated in stop flow mode and fractions were pyrolyzed to allow chromatographic separation of the fragments of the otherwise non-volatile stabilizer. After identification on the basis of accurate mass and elemental composition of the additive the quantification was compared using the available SEC-UV and SEC-PY-GC-TOFMS data. The resulting method provided a high degree of certainty in identification and flexibility in quantification expected to be applicable to other additives of similar volatilities or functional class.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.chroma.2014.08.031DOI Listing

Publication Analysis

Top Keywords

identification quantification
8
quantification high
8
high molecular
8
molecular weight
8
weight light
8
light stabilizer
8
size exclusion
8
flow mode
8
identification
4
stabilizer polycarbonate
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!