Background: A sensitive, specific and robust liquid chromatography-tandem mass spectrometry method has been developed and validated for the quantification of rilpivirine in human plasma, genital/rectal biofluids and mucosal tissues.

Methods: Plasma and tissue samples were extracted using protein precipitation (acetonitrile/water; 5:1 v/v), and genital/rectal biofluids absorbed onto ophthalmic swabs were extracted using liquid-liquid extraction (hexane/ethyl acetate; 80:20 v/v). A stable isotope-labeled internal standard ((13)C-d4-RPV) was used, and the assay was validated over a concentration range of 0.5-400 ng/ml.

Conclusion: Inter- and intra-assay precision and accuracy met the acceptance as per US FDA bioanalytical guidelines. The validated assay has been used for the determination of rilpivirine concentrations in these matrices as part of an exploratory pharmacokinetic study investigating the suitability of a long-acting formulation of rilpivirine for pre-exposure prophylaxis.

Download full-text PDF

Source
http://dx.doi.org/10.4155/bio.14.59DOI Listing

Publication Analysis

Top Keywords

quantification rilpivirine
8
rilpivirine human
8
human plasma
8
liquid chromatography-tandem
8
chromatography-tandem mass
8
mass spectrometry
8
genital/rectal biofluids
8
plasma cervicovaginal
4
cervicovaginal fluid
4
fluid rectal
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!