Dose heterogeneity correction for low-energy brachytherapy sources using dual-energy CT images.

Phys Med Biol

Medical Biophysics Department, University of Toronto, 2075 Bayview Avenue, Toronto, ON M4N3M5, Canada.

Published: September 2014

Permanent seed implant brachytherapy is currently used for adjuvant radiotherapy of early stage prostate and breast cancer patients. The current standard for calculation of dose around brachytherapy sources is based on the AAPM TG-43 formalism, which generates the dose in a homogeneous water medium. Recently, AAPM TG-186 emphasized the importance of accounting for tissue heterogeneities. We have previously reported on a methodology where the absorbed dose in tissue can be obtained by multiplying the dose, calculated by the TG-43 formalism, by an inhomogeneity correction factor (ICF). In this work we make use of dual energy CT (DECT) images to extract ICF parameters. The advantage of DECT over conventional CT is that it eliminates the need for tissue segmentation as well as assignment of population based atomic compositions. DECT images of a heterogeneous phantom were acquired and the dose was calculated using both TG-43 and TG-43 [Formula: see text] formalisms. The results were compared to experimental measurements using Gafchromic films in the mid-plane of the phantom. For a seed implant configuration of 8 seeds spaced 1.5 cm apart in a cubic structure, the gamma passing score for 2%/2 mm criteria improved from 40.8% to 90.5% when ICF was applied to TG-43 dose distributions.

Download full-text PDF

Source
http://dx.doi.org/10.1088/0031-9155/59/18/5305DOI Listing

Publication Analysis

Top Keywords

brachytherapy sources
8
seed implant
8
tg-43 formalism
8
dose calculated
8
calculated tg-43
8
dect images
8
dose
7
tg-43
5
dose heterogeneity
4
heterogeneity correction
4

Similar Publications

Current availability of radiotherapy devices in Peru and artificial intelligence-based analysis for constructing a nationwide implementation plan.

Radiother Oncol

January 2025

Department of Radiotherapy, Instituto Nacional de Enfermedades Neoplasicas, Lima, Peru; Department of Radiation Oncology, Oncosalud - Auna, Lima, Peru. Electronic address:

Purpose: We provide for the first time a comprehensive situational diagnosis and propose an artificial intelligence (AI)-assisted nationwide plan of implementation, attending the most urgent needs.

Methods: Baseline information was collected from open-source databases of the Peruvian Government. Data on cancer incidence from the Health Authorities and GLOBOCAN were collected and compared.

View Article and Find Full Text PDF

Background: High-dose-rate (HDR) brachytherapy using Iridium-192 as a radiation source is widely employed in cancer treatment to deliver concentrated radiation doses while minimizing normal tissue exposure. In this treatment, the precision with which the sealed radioisotope source is delivered significantly impacts clinical outcomes.

Purpose: This study aims to evaluate the feasibility of a new four-dimensional (4D) in vivo source tracking and treatment verification system for HDR brachytherapy using a patient-specific approach.

View Article and Find Full Text PDF

Background: The purpose of this study is to determine the effect of the type of I-125 radioactive source on dose distribution in the planning process of ultra-low dose rate (uLDR) prostate brachytherapy.

Material And Methods: 7 patients who had undergone brachytherapy in our center were included in the study. Dose in five geometrical points were analyzed for 12 types of implants that are available on the market.

View Article and Find Full Text PDF

Optimized chelator and nanoparticle strategies for high-activity Pd-loaded biodegradable brachytherapy seeds.

EJNMMI Radiopharm Chem

December 2024

The Hevesy Laboratory, DTU Health Technology, Frederiksborgvej 399, 4000, Roskilde, Denmark.

Background: Brachytherapy (BT) is routinely used in the treatment of various cancers. Current BT relies on the placement of large sources of radioactivity at the tumor site, requiring applicators that may cause local traumas and lesions. Further, they suffer from inflexibility in where they can be placed and some sources reside permanently in the body, causing potential long-term discomfort.

View Article and Find Full Text PDF

Background: Cervical cancer is the second most common cancer among women in Nigeria where, the gap between need for, and access to, radiation therapy including brachytherapy is significant. This report documents the implementation of the first three-dimensional high-dose-rate (3D-HDR) brachytherapy service for cervical cancer in Nigeria.

Purpose: This report details the steps taken to implement the 3D-HDR brachytherapy program, the challenges faced, and the adaptive strategies employed to overcome them.

View Article and Find Full Text PDF