Background: The purpose of this study was to examine factors that could help reduce primary perforation during insertion of a framed intrauterine device (IUD) and to determine factors that contribute in generating enough uterine muscle force to cause embedment and secondary perforation of an IUD. The objective was also to evaluate the main underlying mechanism of IUD expulsion.
Methods: We compared known IUD insertion forces for "framed" devices with known perforation forces in vitro (hysterectomy specimens) and known IUD removal forces and calculated a range of possible intrauterine forces using pressure and surface area. These were compared with known perforation forces.
Results: IUD insertion forces range from 1.5 N to 6.5 N. Removal forces range from 1 N to 5.8 N and fracture forces from 8.7 N to 30 N depending upon device. Measured perforation forces are from 20 N to 54 N, and calculations show the uterus is capable of generating up to 50 N of myometrial force depending on internal pressure and surface area.
Conclusion: Primary perforation with conventional framed IUDs may occur if the insertion pressure exceeds the perforation resistance of the uterine fundus. This is more likely to occur if the front end of the inserter/IUD is narrow, the passage through the cervix is difficult, and the procedure is complex. IUD embedment and secondary perforation and IUD expulsion may be due to imbalance between the size of the IUD and that of the uterine cavity, causing production of asymmetrical uterine forces. The uterine muscle seems capable of generating enough force to cause an IUD to perforate the myometrium provided it is applied asymmetrically. A physical theory for IUD expulsion and secondary IUD perforation is given.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4132253 | PMC |
http://dx.doi.org/10.2147/IJWH.S63167 | DOI Listing |
Acta Obstet Gynecol Scand
January 2025
Department of Gynecology and Obstetrics, Federal University of Goiás (UFG), Goiânia, Brazil.
Introduction: Intrauterine devices (IUDs) are highly effective contraceptives. Despite their effectiveness, pregnancies can occur during IUD use, and the management of such cases, particularly when the pregnancy is desired, remains controversial.
Material And Methods: We conducted a systematic review and meta-analysis to evaluate outcomes in women who unintentionally conceived while using IUDs and chose to continue their pregnancies.
Cureus
December 2024
Obstetrics and Gynaecology, Khyber Teaching Hospital, Peshawar, PAK.
Background Hysteroscopy, a minimally invasive procedure for diagnosing and treating intrauterine pathologies, can be challenging due to inadequate cervical dilation, leading to procedural difficulties and patient discomfort. Misoprostol, a synthetic prostaglandin E1 analog, is increasingly used for cervical ripening to ease hysteroscopic procedures. Objective To evaluate the efficacy and safety of misoprostol for cervical ripening prior to hysteroscopy.
View Article and Find Full Text PDFJ Assist Reprod Genet
January 2025
Evidence Based Medicine Center, School of Basic Medical Sciences, Lanzhou University, Lanzhou, China.
Poor ovarian response (POR) significantly impacts the success of assisted reproductive technology (ART), and growth hormone (GH) has been proposed as an adjuvant treatment to improve outcomes in POR patients undergoing in vitro fertilization (IVF). A systematic review and meta-analysis were conducted to evaluate the effectiveness of GH in enhancing pregnancy outcomes, registering a protocol on PROSPERO and searching multiple databases up to September 2023. Twelve systematic reviews/meta-analysis and 20 randomized controlled trials (RCTs) involving 1984 patients were included.
View Article and Find Full Text PDFCurr Pharm Des
January 2025
Metabolic Syndrome Research Center, Mashhad University of Medical Sciences, Mashhad, Iran.
Intrauterine Adhesions (IUAs) are characterized by endometrial damage due to endometritis or curettage. Currently, the gold standard for IUA treatment is hysteroscopy, which enables the dissolution of IUA through mechanical or electrosurgical energy. Common strategies to prevent recurrence include the insertion of a balloon catheter or IUD in the uterus.
View Article and Find Full Text PDFFront Med (Lausanne)
January 2025
Department of Gynecology and Obstetrics, International Peace Maternity and Child Health Hospital Shanghai Jiao Tong University School of Medicine, Shanghai Key Laboratory of Embryo Original Diseases, Shanghai Municipal Key Clinical Specialty, Shanghai, China.
Background: An intrauterine device (IUD) is a widely used long-term contraceptive device for family planning. However, the IUD can lead to various complications. Severe complications and remedial measures caused by IUDs have been reported in the literature; however, detailed surgical approaches for safely removing the IUD within the minimum surgical range have rarely been described especially in postmenopausal women.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!