Overview of the repeatability, reproducibility, and agreement of the biometry values provided by various ophthalmic devices.

Am J Ophthalmol

Department of Ophthalmology, Antwerp University Hospital, Edegem, Belgium; Department of Medicine and Health Science, University of Antwerp, Wilrijk, Belgium.

Published: December 2014

Purpose: To present an overview of the measurement errors for various biometric devices, as well as a meta-analysis of the agreement between biometric devices using the Pentacam, Orbscan, and IOL Master as a reference.

Design: Meta-analysis of the literature.

Methods: The meta-analysis is based on data from 216 articles that compare a total of 24 different devices with the reference devices for the following 9 parameters: mean, steep and flat curvature of the anterior and posterior cornea; central corneal thickness; anterior chamber depth; and axial length. After the weighted average difference between devices has been determined, the "two one-sided t test" was used to test for equivalence between devices within certain thresholds defined by the measurement errors and the influence of these differences on the calculated refraction.

Results: In only 17 of the 70 comparisons a device was equivalent with the reference device within the thresholds set by the measurement error. More lenient thresholds, based on a change in calculated refraction of ±0.25 diopter, increased this number to a maximum of 25/50 comparisons (excluding pachymetry). High degrees of inconsistency were seen in the reported results, which could partially explain the low agreement between devices.

Conclusion: As a rule, biometry measurements taken by different devices should not be considered equivalent, although several exceptions could be identified. We therefore recommend that clinical studies involving multiple device types treat this as a within-subject variable to avoid bias. The follow-up of individual patients using different devices should be avoided at all times.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ajo.2014.08.014DOI Listing

Publication Analysis

Top Keywords

devices
9
measurement errors
8
biometric devices
8
overview repeatability
4
repeatability reproducibility
4
reproducibility agreement
4
agreement biometry
4
biometry values
4
values provided
4
provided ophthalmic
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!