In vitro performance and principles of anti-siphoning devices.

Acta Neurochir (Wien)

Department of Neurosurgery, Universitätsmedizin Göttingen, Georg-August University Göttingen, Göttingen, Germany.

Published: November 2014

Background: Anti-siphon devices (ASDs) of various working principles were developed to overcome overdrainage-related complications associated with ventriculoperitoneal shunting.

Objective: We aimed to provide comparative data on the pressure and flow characteristics of six different types of ASDs (gravity-assisted, membrane-controlled, and flow-regulated) in order to achieve a better understanding of these devices and their potential clinical application.

Methods: We analyzed three gravity-dependent ASDs (ShuntAssistant [SA], Miethke; Gravity Compensating Accessory [GCA], Integra; SiphonX [SX], Sophysa), two membrane-controlled ASDs (Anti-Siphon Device [IASD], Integra; Delta Chamber [DC], Medtronic), and one flow-regulated ASD (SiphonGuard [SG], Codman). Defined pressure conditions within a simulated shunt system were generated (differential pressure 10-80 cmH2O), and the specific flow and pressure characteristics were measured. In addition, the gravity-dependent ASDs were measured in defined spatial positions (0-90°).

Results: The flow characteristics of the three gravity-assisted ASDs were largely dependent upon differential pressure and on their spatial position. All three devices were able to reduce the siphoning effect, but each to a different extent (flow at inflow pressure: 10 cmH2O, siphoning -20 cmH2O at 0°/90°: SA, 7.1 ± 1.2*/2.3 ±  0.5* ml/min; GCA, 10.5 ± 0.8/3.4 ± 0.4* ml/min; SX, 9.5 ± 1.2*/4.7 ± 1.9* ml/min, compared to control, 11.1 ± 0.4 ml/min [*p < 0.05]). The flow characteristics of the remaining ASDs were primarily dependent upon the inflow pressure effect (flow at 10 cmH2O, siphoning 0 cmH2O/ siphoning -20cmH2O: DC, 2.6 ± 0.1/ 4 ± 0.3* ml/min; IASD, 2.5 ± 0.2/ 0.8 ± 0.4* ml/min; SG, 0.8 ± 0.2*/ 0.2 ± 0.1* ml/min [*p < 0.05 vs. control, respectively]).

Conclusion: The tested ASDs were able to control the siphoning effect within a simulated shunt system to differing degrees. Future comparative trials are needed to determine the type of device that is superior for clinical application.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00701-014-2201-yDOI Listing

Publication Analysis

Top Keywords

flow characteristics
8
gravity-dependent asds
8
differential pressure
8
asds
6
pressure
6
vitro performance
4
performance principles
4
principles anti-siphoning
4
devices
4
anti-siphoning devices
4

Similar Publications

Impact of adipocytes on ultrasound evaluation of parathyroid adenomas.

J Med Ultrason (2001)

December 2024

Department of Internal Medicine, Kuma Hospital, Kobe, Hyogo, 650-0011, Japan.

Purpose: Parathyroid lipoadenomas are difficult to recognize preoperatively; hence, they may remain undetected. Difficulty in recognition is thought to be due to the adipocytes present in the tumor. This study aimed to clarify the impact of adipocytes as a component of parathyroid adenomas on ultrasound evaluation.

View Article and Find Full Text PDF

Correlation between ultrasonography and elastography parameters and molecular subtypes of breast cancer in young women.

Ann Med

December 2025

Department of Ultrasonographl, Shanxi Bethune Hospital, Shanxi Academy of Medical Sciences, Third Hospital of Shanxi Medical University, Tongji Shanxi Hospital, Taiyuan, Shanxi Province, China.

Objective: To explore the differences of conventional ultrasound characteristics, elastic imaging parameters and clinicopathological characteristics of distinct molecular subtypes of breast cancer in young women, and to identify imaging parameters that exhibited significant associations with each molecular subtype.

Methods: We performed a retrospective analysis encompassing 310 young women with breast cancer. Observations were made regarding the ultrasonography and elastography characteristics of the identified breast lesions.

View Article and Find Full Text PDF

Purpose: This study investigates the capabilities of ultrasonography (US) in determing the stage of orbital inflammation in patients with granulomatosis with polyangiitis (GPA).

Material And Methods: The study included 24 patients (8 men and 16 women) with diffuse orbital tissue involvement in GPA. Group 1 (active stage) included nine patients, while group 2 (inactive stage) consisted of 18 patients.

View Article and Find Full Text PDF

The correlation between Fischer's ratio and the risk of cardiac dysfunction in heart failure patients.

BMC Cardiovasc Disord

December 2024

Department of Cardiovascular Medicine, Rui Jin Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai, 200025, People's Republic of China.

Backgrounds: Due to the high mortality and hospitalization rate in chronic heart failure (HF), it is of great significance to study myocardial nutrition conditions. Amino acids (AAs) are essential nutrient metabolites for cell development and survival. This study aims to investigate the associations and prognostic value of plasma branched-chain amino acid/aromatic amino acid ratio (Fischer's ratio, FR) in patients with left ventricular ejection fraction (LVEF) ≤ 50%.

View Article and Find Full Text PDF

miR-432-5p Targeting SORT1 to Protect Artery Smooth Muscle Cells and Inhibit Coronary Artery Disease.

Biochem Genet

December 2024

Department of Cardiovascular Medicine, Shanghai Baoshan Luodian Hospital, No. 88, Yongshun Road, Baoshan District, Shanghai, 201908, China.

Recent studies highlight the crucial role of microRNAs (miRNAs) in coronary artery disease (CAD). This retrospective study investigated the abundance of miR-432-5p in the serum of CAD patients and explored its role. 252 volunteers were included.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!