Validation of Method Performance of pH, PCO2, PO2, Na(+), K(+) of Cobas b121 ABG Analyser.

J Clin Diagn Res

Tutor, Department of Biochemistry, Pondicherry Institute of Medical Sciences, Affiliated to Pondicherry University, Ganapathichettikulam, Kalapet, Pondicherry, India .

Published: June 2014

Introduction: The introduction of a new method or new analyser is a common occurrence in clinical biochemistry laboratory. Blood gas measurements and electrolytes are often performed in Point-of-Care (POC) settings. When a new POC analyser is obtained, the performance of the analyser should be evaluated by comparison to the measurements with the reference analyser in the laboratory.

Objectives: Evaluation of method performance of pH, PCO2, PO2, Na(+), K(+) of cobas b121 ABG analyser.

Materials & Methods: The evaluation of method performance of pH, PO2, PCO2, Na(+), K(+) of cobas b121 ABG analyser was done by comparing the results of 50 patient samples run on cobas b121 with the results obtained from Rapid lab values (reference analyser). Correlation coefficient was calculated from the results obtained from both the analysers. Precision was calculated by running biorad ABG control samples.

Results: The correlation coefficient values obtained for parameters were close to 1.0 indicating good correlation. The CV obtained for all the parameters were less than 5 indicating good precision.

Conclusion: The new ABG analyser, Cobas b121 correlated well with the reference ABG analyser (Rapid Lab) and could be used to run on patient samples.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4129322PMC
http://dx.doi.org/10.7860/JCDR/2014/9218.4442DOI Listing

Publication Analysis

Top Keywords

cobas b121
20
abg analyser
16
method performance
12
na+ cobas
12
b121 abg
12
analyser
9
performance pco2
8
pco2 po2
8
po2 na+
8
reference analyser
8

Similar Publications

Background: Packed cell volume (PCV) is important for assessing a patient's health status. Some blood gas analyzers measure hematocrit, and the agreement with PCV varies among different analyzers.

Objectives: We aimed to determine the agreement between PCV measured by microcentrifugation and hematocrit measured by the Cobas b121 blood gas analyzer in dogs, cats, and horses.

View Article and Find Full Text PDF

Validation of Method Performance of pH, PCO2, PO2, Na(+), K(+) of Cobas b121 ABG Analyser.

J Clin Diagn Res

June 2014

Tutor, Department of Biochemistry, Pondicherry Institute of Medical Sciences, Affiliated to Pondicherry University, Ganapathichettikulam, Kalapet, Pondicherry, India .

Introduction: The introduction of a new method or new analyser is a common occurrence in clinical biochemistry laboratory. Blood gas measurements and electrolytes are often performed in Point-of-Care (POC) settings. When a new POC analyser is obtained, the performance of the analyser should be evaluated by comparison to the measurements with the reference analyser in the laboratory.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!