Introduction: Follow-up CT angiography (CTA) is routinely performed for post-procedure management after carotid artery stenting (CAS). However, the stent lumen tends to be underestimated because of stent artifacts on CTA reconstructed with the filtered back projection (FBP) technique. We assessed the utility of new iterative reconstruction techniques, such as adaptive statistical iterative reconstruction (ASIR) and model-based iterative reconstruction (MBIR), for CTA after CAS in comparison with FBP.
Methods: In a phantom study, we evaluated the differences among the three reconstruction techniques with regard to the relationship between the stent luminal diameter and the degree of underestimation of stent luminal diameter. In a clinical study, 34 patients who underwent follow-up CTA after CAS were included. We compared the stent luminal diameters among FBP, ASIR, and MBIR, and performed visual assessment of low attenuation area (LAA) in the stent lumen using a three-point scale.
Results: In the phantom study, stent luminal diameter was increasingly underestimated as luminal diameter became smaller in all CTA images. Stent luminal diameter was larger with MBIR than with the other reconstruction techniques. Similarly, in the clinical study, stent luminal diameter was larger with MBIR than with the other reconstruction techniques. LAA detectability scores of MBIR were greater than or equal to those of FBP and ASIR in all cases.
Conclusion: MBIR improved the accuracy of assessment of stent luminal diameter and LAA detectability in the stent lumen when compared with FBP and ASIR. We conclude that MBIR is a useful reconstruction technique for CTA after CAS.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1007/s00234-014-1415-2 | DOI Listing |
J Mol Cell Cardiol
December 2024
A.I.Virtanen Institute for Molecular Sciences, University of Eastern Finland, Kuopio, Finland; Heart Centre and Gene Therapy Unit, Kuopio University Hospital, Kuopio, Finland. Electronic address:
Background: Coronary stenting operations have become the main option for the treatment of coronary heart disease. Vessel recovery after stenting has emerged as a critical factor in reducing possible complications. In this study, we evaluated the feasibility, safety and efficacy of locally administered intraluminal gene therapy delivered using a specialized infusion balloon catheter.
View Article and Find Full Text PDFVasc Endovascular Surg
December 2024
Baylor Scott and White Heart and Vascular Hospital, Dallas, TX, USA.
: Contemporary treatments of acute aortic dissection, including medical, surgical, and endovascular options, are remarkably effective at the management of malperfusion and rupture. Unfortunately, long-term studies indicate that 30%-50% of patients need secondary procedures to treat progressive aneurysmal enlargement of the untreated aorta. The Stent-assisted balloon-induced intimal disruption and relamination in aortic dissection repair (STABILISE) technique was introduced to improve long-term outcomes.
View Article and Find Full Text PDFJ Surg Case Rep
January 2025
Department of Surgery, Mercy University Hospital, Grenville Place, Cork, T12 WE28, Ireland.
Endoscopic management of transmural oesophageal defects following esophagectomy or spontaneous perforations, such as Boerhaave's syndrome, is often complicated by stent migration and luminal occlusion [1]. The Vacuum-Assisted Closure (VAC) stent, which integrates a covered stent with endoscopic vacuum therapy, aims to address these issues by providing functional drainage and promoting wound healing [2]. This case series presents our initial experience with VACStent therapy in four patients treated between February 2023 and April 2024.
View Article and Find Full Text PDFEuroIntervention
December 2024
St. Francis Hospital, Roslyn, NY, USA.
Am J Transplant
November 2024
Isla Technologies, Inc, San Carlos, California, USA. Electronic address:
In this study using a discordant, xenogeneic, transplant model we demonstrate the functionality and safety of the first stent-based bioartificial pancreas (BAP) device implanted endovascularly into an artery, harnessing the high oxygen content in blood to support islet viability. The device is a self-expanding nitinol stent that is coated with a bilayer of polytetrafluoroethylene that forms channels to hold islets embedded in a hydrogel. We completed a 1-month study in the nondiabetic swine model (N = 3) to test the safety of the device and to assess islet functionality after device recovery.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!