Use of multiplex PCR in diagnosis of bloodstream infections in kidney patients.

Diagn Microbiol Infect Dis

Servicio de Microbiología, Hospital Universitario La Paz, IdiPAZ, Paseo de la Castellana, 261, 28046, Madrid, Spain. Electronic address:

Published: October 2014

The LightCycler® SeptiFast Test (Roche Diagnostics GmbH, Mannheim, Germany) was prospectively compared with the standard blood culture technique in a series of 86 kidney patients. The sensitivity of the PCR compared with the culture was 71%, and the specificity was 88%. All the species identified by culture in these patients were in the SeptiFast panel. The median time to results was 1 day for the PCR, 3 days for positive cultures, and 5 days for negative cultures.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.diagmicrobio.2014.07.001DOI Listing